OM0284
Page 63 of 72
Rev 2 July/2021
10.1.4 Maintenance
The output of an ultraviolet light deteriorates with burning age. The useful life of the light is approximately 5000
hours under specific test conditions.
Note:
Before testing with lamp off, the light may be cleaned with a lint-free cloth dampened with alcohol
or ammonia and water.
The control system monitors UV light usage hour and when the 5000 hour point is reached. The UV light LED
indicator will blink fast along with the display message “Replace
UV light”. To view hours or reset timer, perform
the following:
•
Press and hold ENT key for 3 seconds until the menu appears.
•
Press ↑ or ↓ keys to select UV lamp hours.
•
Press ENT key to highlight hours.
•
Press ↑ or ↓ key
simultaneous until hours zero.
•
Press ENT key to accept cleared hours.
•
Press
key to return to run display.
Main Menu
Time Menu
UV Lamp Hours
3356
Temperature 24.0
Blower On/Off Password none
Содержание LabGard NU-543-300E
Страница 7: ...OM0284 Page 7 of 72 Rev 2 July 2021...
Страница 9: ...OM0284 Page 9 of 72 Rev 2 July 2021...
Страница 10: ...OM0284 Page 10 of 72 Rev 2 July 2021...
Страница 11: ...OM0284 Page 11 of 72 Rev 2 July 2021...
Страница 12: ...OM0284 Page 12 of 72 Rev 2 July 2021...
Страница 16: ...OM0284 Page 16 of 72 Rev 2 July 2021...
Страница 21: ...OM0284 Page 21 of 72 Rev 2 July 2021 Figure 1...
Страница 22: ...OM0284 Page 22 of 72 Rev 2 July 2021...
Страница 35: ...OM0284 Page 35 of 72 Rev 2 July 2021...
Страница 67: ...OM0284 Page 67 of 72 Rev 2 July 2021...
Страница 68: ...OM0284 Page 68 of 72 Rev 2 July 2021...
Страница 69: ...OM0284 Page 69 of 72 Rev 2 July 2021...
Страница 70: ...OM0284 Page 70 of 72 Rev 2 July 2021...
Страница 71: ...OM0284 Page 71 of 72 Rev 2 July 2021...
Страница 72: ...OM0284 Page 72 of 72 Rev 2 July 2021...