7.5.4.5 Nite Care Calibration
The Nite Care mode is defaulted to operate the blower at approximately 600 rpm or a 14% duty cycle.
However, if desired the Nite Care blower speed can be adjusted higher or lower by performing the following:
•
Press and hold
key, then press ENT
↑ and ↓ keys sequentially releasing the
key after the 3
key sequence.
The following service parameter display should appear
•
Press ↑ and ↓ to select calibration
•
Select and press ENT key for blower speed calibration
•
Select and press ENT key for Nite Care
Adjust blower speed with ↑ (increase) and ↓ (decrease) as needed.
•
Press ENT key to accept Nite Care blower speed calibration
Service Parameters
Configuration
Setpoints
Calibration
Options
Nite Care Blower Configuration
.41 600 rpm, 14.0%
fpm, 26.0⁰
Settling time
Содержание LabGard NU-543-300E
Страница 7: ...OM0284 Page 7 of 72 Rev 2 July 2021...
Страница 9: ...OM0284 Page 9 of 72 Rev 2 July 2021...
Страница 10: ...OM0284 Page 10 of 72 Rev 2 July 2021...
Страница 11: ...OM0284 Page 11 of 72 Rev 2 July 2021...
Страница 12: ...OM0284 Page 12 of 72 Rev 2 July 2021...
Страница 16: ...OM0284 Page 16 of 72 Rev 2 July 2021...
Страница 21: ...OM0284 Page 21 of 72 Rev 2 July 2021 Figure 1...
Страница 22: ...OM0284 Page 22 of 72 Rev 2 July 2021...
Страница 35: ...OM0284 Page 35 of 72 Rev 2 July 2021...
Страница 67: ...OM0284 Page 67 of 72 Rev 2 July 2021...
Страница 68: ...OM0284 Page 68 of 72 Rev 2 July 2021...
Страница 69: ...OM0284 Page 69 of 72 Rev 2 July 2021...
Страница 70: ...OM0284 Page 70 of 72 Rev 2 July 2021...
Страница 71: ...OM0284 Page 71 of 72 Rev 2 July 2021...
Страница 72: ...OM0284 Page 72 of 72 Rev 2 July 2021...