OM0280
Page 35 of 63
Rev 3 July/2021
7.4 Sliding Window Replacement and Manual Adjustment
The sliding window replacement is accomplished by removing the front decorative panel, control center, and window glide
assemblies. The sliding window adjustment may be required due to everyday use over the life of the cabinet. Both window
glides are adjustable by a set screw and tension screw method. When adjusting the sliding window, be sure to verify
proper micro switch operation. If the sliding window is too loose, it will not properly activate the micro switches, thus
causing potential operational malfunctions to occur. If the window counter-balance experiences a fault, window function
must be verified (the sliding window retention or ability to slow the rate of fall). See Sliding Window Retention Verification
in the Inspection Report).
7.5 Airflow Control System Setup and Calibration
7.5.1
General
The operation of the NU-540E cabinet requires that the setup and calibration procedures be performed in order to
certify or commission the cabinet for usage. The setup and calibration procedures performed
ONLY BY THE
CABINET CERTIFIER
ensure that cabinet's setpoints are verified and that the airflow monitor sensor is calibrated to
the correct values.
Содержание LabGard NU-540-300E
Страница 7: ...OM0280 Page 7 of 63 Rev 3 July 2021...
Страница 9: ...OM0280 Page 9 of 63 Rev 3 July 2021...
Страница 10: ...OM0280 Page 10 of 63 Rev 3 July 2021...
Страница 11: ...OM0280 Page 11 of 63 Rev 3 July 2021...
Страница 12: ...OM0280 Page 12 of 63 Rev 3 July 2021...
Страница 16: ...OM0280 Page 16 of 63 Rev 3 July 2021...
Страница 20: ...OM0280 Page 20 of 63 Rev 3 July 2021...
Страница 58: ...OM0280 Page 58 of 63 Rev 3 July 2021...
Страница 59: ...OM0280 Page 59 of 63 Rev 3 July 2021...
Страница 60: ...OM0280 Page 60 of 63 Rev 3 July 2021...
Страница 61: ...OM0280 Page 61 of 63 Rev 3 July 2021...
Страница 62: ...OM0280 Page 62 of 63 Rev 3 July 2021...
Страница 63: ...OM0280 Page 63 of 63 Rev 3 July 2021...