OM0280
Page 56 of 63
Rev 3 July/2021
11.0 Electrical/Environmental Requirements
11.1 Electrical (Supply Voltage Fluctuations Not to /- 10%)
*NU-540-300E
230 VAC,
50/60 Hz,
1 Phase,
8 Amps
*NU-540-400E
230 VAC,
50/60 Hz,
1 Phase,
10 Amps
*NU-540-500E
230 VAC,
50/60 Hz,
1 Phase, 11 Amps
*NU-540-600E
230 VAC,
50/60 Hz,
1 Phase, 11 Amps
11.2 Operational Performance (for indoor use only)
Environment Temperature Range:
60
F-85
F (15
C - 30
C)
Environment Humidity:
Maximum relative humidity 80% for temperatures up to 31
C decreasing linearly to
50% relative humidity at 40
C.
Environment Altitude:
6562 Feet (2000M)
11.3 Light Exposure
Standard Fluorescent Lighting @ 150 ft. candles (1614 LUX) maximum intensity.
11.4 Installation Category: 2.0
Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to
withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example,
in CAT II, which is the category used for instruments in installations supplied from a supply comparable to public mains
such as hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500 V
for a 230 V supply and 1500 V for a 120 V supply.
11.5 Pollution Degree: 2.0
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2
assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity
caused by condensation.
11.6 Chemical Exposure
Chemical exposure should be limited to antibacterial materials used for cleaning and disinfecting.
USE OF CHLORINATED
OR HALOGEN MATERIALS IN THE CABINET MAY DAMAGE STAINLESS STEEL
. Equipment decontamination can be
accomplished by non-condensing gas or vapor paraformaldehyde, Hydrogen Peroxide or Chlorine Dioxide following
NSF/ANSI Annex G or EN 12459 Annex J.
11.7 EMC Performance (classified for light industrial)
Emissions:
EN61326
Immunity:
EN61326
Class A equipment is intended for use in an industrial environment.
In the documentation for the user, a statement shall be included drawing
attention to the fact that there may be potential difficulties in ensuring
electromagnetic compatibility in other environments, due to conducted
as well as radiated disturbances.
WARNING
!
Содержание LabGard NU-540-300E
Страница 7: ...OM0280 Page 7 of 63 Rev 3 July 2021...
Страница 9: ...OM0280 Page 9 of 63 Rev 3 July 2021...
Страница 10: ...OM0280 Page 10 of 63 Rev 3 July 2021...
Страница 11: ...OM0280 Page 11 of 63 Rev 3 July 2021...
Страница 12: ...OM0280 Page 12 of 63 Rev 3 July 2021...
Страница 16: ...OM0280 Page 16 of 63 Rev 3 July 2021...
Страница 20: ...OM0280 Page 20 of 63 Rev 3 July 2021...
Страница 58: ...OM0280 Page 58 of 63 Rev 3 July 2021...
Страница 59: ...OM0280 Page 59 of 63 Rev 3 July 2021...
Страница 60: ...OM0280 Page 60 of 63 Rev 3 July 2021...
Страница 61: ...OM0280 Page 61 of 63 Rev 3 July 2021...
Страница 62: ...OM0280 Page 62 of 63 Rev 3 July 2021...
Страница 63: ...OM0280 Page 63 of 63 Rev 3 July 2021...