NOVUS AUTOMATION
22/57
•
Debounce:
If the sensor type configured is Dry Contact, it is necessary to set a
debounce
time for edge detection. The
debounce
is the sensor
stabilization time - the minimum time at which the sensor must remain at the logical level of interest so that the detected edge is considered
valid. The minimum configurable
debounce
time is 50 milliseconds and the maximum are 6 seconds.
•
User Unit:
Allows you to configure the flow unit related to the pulses counted in the Digital Input.
LogBox-BLE
provides 9 flow units. If one of
them is selected, simply configure the sensor factor together with the sensor unit, so that the device calculates the counts and displays the flow
rate already on the configured unit. If none of the units provided meets the need, you can select the Custom unit.
o
Custom:
Allows you to configure a custom unit for the digital input. In this
mode it is possible to describe the unit with up to 8 characters. If
you configure a custom unit, you need to configure the User Factor.
•
No. of Decimal Places:
Allows you to configure the desired number of decimal places to display the calculated value, in flow, of the digital
input.
•
Sensor Factor:
Allows you to configure the sensor factor used at the digital input. This parameter can be found in the sensor manual as
k-
factor
.
LogBox-BLE
provides 3 units for the sensor. If none of the units provided meets the need, you can select the Custom unit.
•
User Factor:
If the User Unit or Sensor Factor unit is set to Custom, you need to configure the User Factor. The User Factor must relate the
required unit to the Sensor Unit and the Sensor Factor. It will be used as a factor to be multiplied in the read counts of the digital input. For more
details, refer to chapter INPUT SIGNALS READING.
•
Lower Alarm:
Allows you to enable and configure a lower alarm setpoint.
o
Hysteresis:
Allows you to configure a hysteresis for the lower alarm.
•
Upper Alarm:
Allows you to enable and configure an upper alarm setpoint.
o
Hysteresis:
Allows you to configure a hysteresis for the upper alarm.
EVENT REGISTER OR REGISTERS CONTROL
NXperience
NXperience-BLE
Android
iOS
Fig. 07
– Event Register or Registers Control Mode
•
Tag:
Allows you to configure a 16-character name for the digital channel. This name will be used to identify the channel in a data download.
•
Input Mode:
Allows you to select the digital input mode. In this example, the selected mode must be Event Log or Log Control. If you select the
Log Control mode, in the DATA LOGGING setting, the Start and/or Stop of Logs mode via Digital Input must be selected for the configuration to
take effect.
•
Sensor Type:
Allows you to configure the sensor type to be connected to the digital input: PNP, NPN, or Dry Contact.
•
Event Edge:
Allows you to configure the desired event edge. This way, the device will log events each time the configured edge is detected at
the digital input. In Event Log mode you can configure to log on the Rising, Falling or Both edges. In the Log Control mode, it is possible to
select to control logs in the Rising, Falling, Both edges - Logs at logical level ‘1’ or Both - Logs at logical level ‘0’.