![Novatel GPStation-6 Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/novatel/gpstation-6/gpstation-6_user-manual_1711304071.webp)
71
GPStation-6 User Manual Rev 2
Logs
5.1.7
ISMCALIBRATIONSTATUS TEC self-calibration status
This log shows the status of a TEC self-calibration (see ISMCALIBRATE command) procedure used
to determine TEC bias caused by receiver RF delay. Once the calibration is complete the TEC
calibration values reported in this log can be used with the ISMTECCALIBRATION command to set
the TEC offsets used in the receiver calculations. If this log is output ONNEW it will stop being
produced once the calibration duration set using the ISMCALIBRATE command has passed. It can
also be logged ONCE at the end of calibration to report the final computed values.
The following information is included in this log:
•
Duration that the calibration has been running at the time the log is output
•
Number of measurement samples that have been included in calculating the current TEC
calibration value. Only measurements above the elevation cutoff specified with the
ISMCALIBRATE command are included in the calibration.
•
The current TEC calibration value in units of TECU
•
The standard deviation of the current TEC calibration value
Structure:
Message ID: 1406
Log Type: Asynch
Recommended Input:
LOG
ISMCALIBRATIONSTATUSA
ONNEW
ASCII example:
#ISMCALIBRATIONSTATUSA,USB1,0,81.0,FINESTEERING,1666,430620.000,00000020,45d2
,10283;1680,2,
GPSL1CAL2Y,28,-27.513463974,1.702342391
GLOL1CAL2P,10,7.925333500,1.621402383
Field
Field Type
Data Description
Format
Binary
Bytes
Binary
Offset
1
ISMCALIBRATION
STATUS Header
Log header
H
0
2
Calibration time
Time in calibration mode (seconds)
ULONG 4
H
3
# records
Number of records in this log
ULONG 4
H+4
4
Signal type
Signal combination type (refer to
Table 20,
Signal Combination Type
ENUM
4
H+8
5
# samples
Number of samples collected
ULONG 4
H+12
6
TEC calibration
Current computed TEC calibration value
(TECU)
FLOAT
4
H+16
7
TEC Std.Dev
Standard deviation of TEC calibration (TECU) FLOAT
4
H+20
8 ...
Next record offset =
H + 8 + (# records x
20)
variable
xxxx
32-bit CRC (ASCII and Binary only)
HEX
4
H+8+ (#
records
x 20)
variable
[CR][LF]
Sentence terminator (ASCII only)
-
-
-