8 Program selection
Battery data entry
42
©
Nortec Electronics 2018
10: Pb battery sealed MIL
P410:
Please enter parameter:
RATED CAPACITY ? 000.0 Ah
The nominal capacity is measured in ampere hours (Ah). The energy content of the cell determines
at the maximum admitted current for charging and discharging.
The identification plate of the battery contains the capacity information. Depending on the choice of
the level of the discharging current the batteries of identical construction types have different
capacities.
Rule of thumb: If the chosen discharging current is higher, the capacity available will be lower.
Military lead batteries are often defined according to the “5 hour-capacity”.
The definition of the capacities for civil batteries is based on the area of application and the norm to
be applied differs from 10hours to 100hours discharging currents. The treatments which are
implemented in the UL60 consider these problems. After entering the capacity possible in steps of
0.1
Ah
confirm with
and a start menu will be displayed:
10: Lead battery sealed MIL
Program:
Maintenance
Start:
=with print =without print
Entering
or
you will return to a mask where battery type and treatment can
again be selected.
Entering
starts a testing routine with discharging with 5-hours current as the first
step. Press the button
in order to get a printed record. Enter
battery number
user number
and confirm with
Entering
starts a testing routine after battery number and user number have
been given in and confirmed with
.