Version date: March 27
th
2018
49
7
Operator Manual
NA 3338 Rapid Combi APM
13.Remove
the three
O rings from the pump
foot.
14.Remove
the valve
hose from the pump
foot.
15.Remove
the mix valve
from the pump foot.
cleaning
and
disinfecting
16.Clean
and disinfect all
components and dry
them carefully with a
clean towel (prefera-
bly paper). (see para-
graph 7.2.3)
17.Fit
the valve hose to
the pump foot.
DO NOT
lubricate it!
18.Fit
the mix valve to
the pump foot.
DO NOT
lubricate it!
19.Fit
the 3 O rings to the
pump foot.
DO NOT
lubricate it!
20.Lubricate
all O rings
with Petrogel.
21.Fit
the thick O ring to
the bottom disc.
22.Fit
the O rings to the
outside of the top
disc.
23.Fit
the two O rings to
the outside of the
middle disc.
24.Fit
the O rings to the
inside of the middle
disc.