5.2.71 Error 2201 (Calibration sensor error)
Is the connector of the touch sensor type
calibration jig (40300-0970) connected to the jack
for maintenance on the USB board?
No
Connect the connector of the touch sensor type
calibration jig (40300-0970) to the jack for
maintenance on the USB board
.
Yes
Are there any breaks in the cable of the touch
sensor type calibration jig (40300-0970)?
Yes
Replace the touch sensor type calibration jig
(40300-0970)
.
No
Are there any breaks in the cable ASSY (40274-
CA11)?
Yes
Replace the cable ASSY (40274-CA11)
.
No
Replace the USB board (44501-BA13)
.
Is the problem resolved?
Yes
End
No
Replace the main board (44501-BA01)
.
Содержание SE-9090 Supra
Страница 1: ...SYSTEM EDGER SE 9090 Supra SERVICE MANUAL September 2012 Pages in total 350 LLE7D RDA002B...
Страница 12: ...LLE7D RDA002B X This page is intentionally left blank...
Страница 14: ...LLE7D RDA002B 2 This page is intentionally left blank...
Страница 22: ...LLE7D RDA002B 10 This page is intentionally left blank...
Страница 24: ...LLE7D RDA002B 12 This page is intentionally left blank...
Страница 104: ...LLE7D RDA002B 92 This page is intentionally left blank...
Страница 310: ...LLE7D RDA002B 298 9 3 Wiring Diagram...
Страница 311: ...LLE7D RDA002B 299 9 4 Connector Cable...
Страница 312: ...LLE7D RDA002B 300...
Страница 313: ...LLE7D RDA002B 301 9 5 Layout Drawing 9 5 1 ASSYs 40274 2000...
Страница 314: ...LLE7D RDA002B 302 40274 3000 40274 5000 40274 6000...
Страница 315: ...LLE7D RDA002B 303...
Страница 316: ...LLE7D RDA002B 304 40274 4000 40274 8000...
Страница 317: ...LLE7D RDA002B 305 40274 7100 40274 7200 40273 7200 2 2 5 2 5 5 2 2 2 5...
Страница 318: ...LLE7D RDA002B 306 9 5 2 Boards S N 700019 and earlier...
Страница 319: ...LLE7D RDA002B 307 S N 700020 and later...
Страница 320: ...LLE7D RDA002B 308 9 5 3 Cable ASSYs...
Страница 321: ...LLE7D RDA002B 309 9 5 4 Other electrical components...
Страница 322: ...LLE7D RDA002B 310...
Страница 350: ...LLE7D RDA002B 338 This page is intentionally left blank...