5
1.2 Checking chuck pressure
Adjustment Parameter:
Press and hold
(
STOP
)
,
then press
to go to Parameter screen. Select
Parameter setting, then press
. Scroll page to
Parameter setting
(9/12 page)
*
Chuck pressure
(
CR39/HPL
):
100
Chuck pressure
(
PC/ACR/TRX/PUR
):
100
Chuck pressure
(
GLS
):
100
1. Call up a JOB. Any job is OK.
2. Set the chuck pressure JIG into the machine, then press
,
then
(
START
)
.
3. When you press
(
START
)
a lens is chucked. Then press
(
STOP
)
to stop the
process. Measure chuck pressure. Proper chuck pressure is the following.
4. Chuck
pressure
(
CR39/HPL
)
:
45kg+/-5kg
Chuck pressure
(
PC/ACR/TRX/PUR
):
45kg+/-5kg
Chuck pressure
(
GLS
):
45kg+/-5kg
If not in tolerance, change parameter value to adjust chuck pressure.
2. Calibration and Adjustment
Adjustment Order
2-1Calibration
2-2 Finish Size and Bevel Position Adjustment
2-3 AXIS Adjustment
2-4 Feeler Axis Adjustment
2-5 Polish Adjustment
2-6 Groove Adjustment
2-7 Safety Bevel Adjustment
2-8 High Curve grinding Adjustment
To call up internal 45 circle : Press and hold
then press
To call up internal 45 square : Press and hold
then press FC SW.
*
LEX1000 new software (V1.05 and later) has 16 menu pages. Menu pages 14~16 are LEX Drill parameters.
Содержание Lex 1000
Страница 1: ...Installation Adjustment Manual Rev 11 18 08...
Страница 2: ......
Страница 25: ...23 Notes...
Страница 26: ...24 Notes...
Страница 27: ...25 Notes...
Страница 28: ...325 Oser Avenue Hauppauge NY 11788 www Santinelli com 800 644 3343 631 435 9200 November 2008...