22
AXIS adjustment (HCrvFr and HcrvRe)
Adjustment Parameter :
Press and hold
(
STOP
)
,
then press
to go to Parameter screen. Select
Parameter setting, then press
. Scroll the page to
Parameter setting
(Page
11/12)
AXIS adjustment (HCrvFr)
:
0.00
AXIS adjustment (HcrvRe)
:
0.00
1. Call up internal 45 square. Press START. After LMU measures, in Bevel Simulation
Screen,
change “TILT” value to 0.0
, then press START.
2.
After the Front Bevel is completed, press STOP. Check axis on the front side.
3.
Checking axis on the front side. View the front side of lens. If tilted in CW direction,
please decrease the value. If tilted in CCW direction please increase the value.
4.
Repeat the process for the rear side.
5.
Checking axis on the rear side. View the rear side of lens. If tilted in CW direction,
please increase the value. If tilted in CCW direction please decrease the value.
View from
Front
+
–
–
+
Rear
View from