7 - 2
Symptom
Suggestion
The printer does operate, however,
printed results cannot be obtained.
・
The printer roll may be set with the wrong side up. Set it
with the correct side up.
"Err" appears.
・
・
・
・
・
The patient may have blinked during measurement. Instruct
the patient not to blink, and try again.
The eyelid or eyelashes may be obstructing the
measurement. Instruct the patient to open the eye widely. If
his/her opening is not enough, open it with your hand,
paying attention not to press against the eye ball.
The pupil may be too small for measurement. Let the
patient be in a dark room for a while until the pupil opens
enough, and try again.
If you are measuring an IOL-implanted eye or an eye
wearing a CL, press
and try again.
IOL appears on the monitor.
The data may have exceeded the measurable limit.
Date and time setting goes wrong, or
the system parameters and comment
settings return to the factory settings.
・
The internal battery may be dead. Turn the power on and
leave it for 24 hours to recharge the battery.
* If the problem cannot be solved by following the suggestions above, contact your
authorized distributor.