6 - 4
{
{
{
{
{
Parameter Table
< Setting Mode 1 >
* The underlined items indicate factory-settings.
[Item No. 1] The indication step of SPH and CYL data for AR measurement.
[Item No. 2] The distance between the corneal vertex to the posterior surface of the spectacle lens
when patient wears it.
[Item No. 3] The indication step of AXIS data for AR measurements.
[Item No. 4] The way of fogging for AR serial measurement.
Con. ....................... By holding down the start button, the fog keeps on throughout
the serial measurement. It is useful for children who do not
fixate their eyes long.
Nor......................... The fog will be released for each measurement even though the
start button is held down. It is useful for patients who accom-
modate one’s sight easily.
[Item No. 5] How to start printing, or not print.
Manu. .................... Data can be printed out by pressing
.
Auto....................... Data will be printed out after measurement is completed.
NO ......................... Data will not be printed out.
[Item No. 6] Selection of whether or not to use the AI mode.
When set to “YES”, the serial measurement completes when the selected value is
obtained. But if “Error Data” of the parameter is set to “YES” and there are not any
variations in the three measurements, measurement ends. If there is an error data (confidence
coefficient=E) in the measurements, the selected value is not obtained.
[Item No. 7] Selection of whether or not to use the Auto IOL function.
When set to “YES”, the IOL mode automatically turns on after three times of erroneous
measurements.
[Item No. 8] Selection of the tone of beeps which sound during measurement.
[Item No. 9] Selection of whether or not to use the economical print function.
When set to “YES”, the printing will come out with narrow line-spacing so that the
printer roll can be saved.
[Item No. 10] The format of printing measurement results. (See “5. PRINTOUT”.)
Short ...................... Only the restricted data will be printed out. (Short print)
User ....................... Items that are specified in parameter item # 11 - 19 will be printed
out.
(High-speed mode)
No. Items
Contents
1
Step
0.01D / 0.12D / 0.25D
2
Vertex D.
0mm / 10.5mm / 12mm /
13.75mm / 15mm / 16.5mm
3 Axis
Step
1° / 5°
4 Meas.
Mode
Con. / Nor.
5
Manu. / Auto / NO
6
AI Mode
YES / NO
7 Auto
IOL
YES / NO
8
Beep
High / Low / NO
9
Econo. Print
YES / NO
10
Print Format
Short / User
Содержание AR-600
Страница 1: ...NIDEK AUTO REFRACTOMETER Model AR 600 OPERATOR S MANUAL...
Страница 5: ...This page is intentionally left blank...
Страница 16: ...2 9 2 9 Labels The following labels provide safety information about each part or or or...
Страница 37: ...6 2 No Item Content 3 In the selected mode press Each content will be printed as follows...
Страница 65: ......