3 - 6
Measuring Window
Patient looks at the chart through this win-
dow.
Eye Level Marker
A guide for the patient’s eye level for mea-
surement. The height of the chin rest should
be adjusted so that the center level of the
patient’s eye almost aligns with this line.
Chin Rest Knob
Moves the chin rest up and down by turning
this. You may use the eye level marker as a
guide to adjust the patient’s eye level to the
comfortable height for measurement.
Fuses
Be sure to use the specified ones.
Power Inlet
The attached power cord is to be connected
here.
Interface Connectors
*
1
side :
Connect the interface cable to transfer mea-
sured results to an external computer etc.
side :
Connect the interface cable from the
lensmeter to import measured results from the
lensmeter. If the AR-600 is equipped with an
IC card system, the connector on
side can
not be used.
By connecting the Lensmeter to
side of the
interface connectors and the refractor RT-2100/
RT-1200 to
side, the LM (Lensmeter) data
can be transmitted to the refractor via the AR.*
2
By pressing the print button on the Lensmeter,
the data transmission will be started.
H Control of TV Monitor
Adjusted with a screwdriver when the display
does not lock horizontally. Can be found under
the main body.
V Control of TV Monitor
Adjusted with a screwdriver when the display
does not lock vertically. Can be found under the
main body.
*1 Accessory equipment connected to the analog and digital interfaces must be certified according
to the representative appropriate national standards (for example, UL 1950 for Data Processing
Equipment UL 2601-1 for Medical Equipment, and CSA C22.2 No. 601-1, EN 60601-1 and
IEC 60601-1.) Further more all configurations shall comply with the system standard IEC
60601-1. Everybody who connects additional equipment to the signal input part or signal
output part configured a medical system, and is therefore responsible that the system complies
with the requirements of the system standard IEC 60601-1. If in doubt, consult the technical
service department or your local representative.
*2 When communicating with the Lensmeter, set communication parameters for each instrument
as follows. For the setting procedure, see the Operator’s Manual of each instrument.
• AR-600
• NIDEK Lensmeter
28: Baud-Rate = 9600
RS-232C = NIDEK
29: Bit Length = 8
Baud-Rate = 9600
Parity = Odd
Data Bits = 8
Stop Bits = 1
Содержание AR-600
Страница 1: ...NIDEK AUTO REFRACTOMETER Model AR 600 OPERATOR S MANUAL...
Страница 5: ...This page is intentionally left blank...
Страница 16: ...2 9 2 9 Labels The following labels provide safety information about each part or or or...
Страница 37: ...6 2 No Item Content 3 In the selected mode press Each content will be printed as follows...
Страница 65: ......