5.  INSTALLATION 

 

 

5.1  SETTING UP THE ILLUMINATOR 

 

Place the illuminator within the low on shelf of the supplied NICO Myriad NOVUS cart. 

NOTE 

If used without the required NOVUS cart, avoid locations where the 
illuminator may be splashed with liquid. 

            DO NOT use in any environment with explosive or flammable gases. 

This illuminator pulls cooling air into its cabinet through a filtered grille in the 
bottom panel underneath the front portion and exhausts warm air from the fan in 
the rear. 

 
 

DO NOT block the space in front or below the front of the illuminator or impede 

 

exhaust air flow behind or above the cabinet. 

 

If used without the required NOVUS cart a clear space minimum of 5”/12.7cm 
behind and above the cabinet is required.  
The illuminator should not be placed where its exhaust will influence other devices, 
nor where exhaust from other devices will influence the illuminator. 

Ensure the rear power switch is in OFF position. 
 
Ensure that the lamp is properly positioned for use.  Open the lamp compartment door.  
Move the lamp positioning lever (found to the right of the lamp) clockwise, from horizontal 
to vertical.  As the positioning cones disengage the lamp should drop about 3mm.  Move 
the lamp positioning lever counterclockwise to reengage the positioning cones and 
observe that the lamp rises about 3mm as it returns to operating position.   
 
Close the lamp compartment door.   
 
Connect AC power cord to the power inlet located on the rear panel of the illuminator.  
Ensure the power cord is completely seated into the power inlet and engage the cord 
retainer if so equipped.  

 

 

CAUTION 

Use only power cords provided with the illuminator. 

Plug the AC power cord into a wall outlet using the three (3) prong plug 
supplied with the unit. 

 

 

CAUTION

 

To prevent electric shock, connect power cords of peripheral equipment 
through medical isolation transformers. 

 

NOTE 

When using medical isolation transformer, be sure to check the transformer 
power ratings. Ensure that the power cord is connected to the main power with 
three prong plug (USA use UL2601-1 rated isolation transformers and/or power 
strips only). 

 

 

 

CAUTION

 

Nearby radio or TV units may subject this equipment to radio interference. 
To avoid adverse electromagnetic effects, DO NOT operate this equipment 
near RF energy equipment.  

Содержание myriad-LX

Страница 1: ...ith Digital Dimming Control Operator Manual DIST NICO Corporation 250 E 96TH Street Suite 125 Indianapolis IN 46240 USA Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville FL 32216 USA EC REP AJ...

Страница 2: ...OTICES 6 2 2 NORMAL OPERATING DISPLAY 6 2 2 1 LAMP HOURS DISPLAY 6 2 2 2 LIGHT BRIGHTNESS CONTROL 6 2 2 3 STANDBY 6 2 2 4 MENU 6 2 2 4 1 RESET FILTER 6 2 2 4 2 ERROR LOG 6 2 2 4 3 TOTAL UNIT HOURS 6 2...

Страница 3: ...ource utilizing state of the art illumination technology It offers a variety of features such as 5600 K color temperature provides daylight brightness for perfect color definition Quiet operation Comp...

Страница 4: ...atements of suitability for any purpose Caution All devices connecting to the Illuminator must be classified as medical equipment Additional information processing equipment connected to the Illuminat...

Страница 5: ...y with the same type and rating of lamp Read instructions before replacing lamp Caution Do not operate illuminator without lamp in place Caution For grounding reliability use only hospital grade marke...

Страница 6: ...50 60 Hz 220 240 V AC 50 Hz Power consumption 750 watt max Regulatory Approvals UL60601 EN 60601 1 EN 60601 1 2 CAN CSA C22 2 No 601 1 M90 Equipment Class BF type Conforms to CISPR 11 Class A Mode of...

Страница 7: ...Panel No Name Function 1 Lamp switch Turns the illuminator on and off illuminates blue when turned on 2 Touch screen Controls light intensity displays notices and accesses data 3 Turret Autoshutter t...

Страница 8: ...t 2 Circuit breaker Over current protection 3 AC mains power switch and input module Accepts AC Power cord and switches on AC power Green light indicates on 4 Product label Product Label containing Mo...

Страница 9: ...door Move the lamp positioning lever found to the right of the lamp clockwise from horizontal to vertical As the positioning cones disengage the lamp should drop about 3mm Move the lamp positioning le...

Страница 10: ...6 OPERATION 6 1 POWERING UP THE ILLUMINATOR To operate the illuminator First attach the fiber to the ILLUMINATOR Turn on the rear power switch Turn on the front lamp switch The lamp switch will light...

Страница 11: ...to assure full engagement to the power connectors Lower the lamp engagement lever counter clockwise from vertical to horizontal and observe that the lamp rises slightly as it is engaged into proper o...

Страница 12: ...ed more than 1000 hours Install new STX0350 lamp or one used less than 1000 hours The main screen always displays lamp hours and notices are displayed when a lamp is approaching the end of its useful...

Страница 13: ...Turn off and restart the illuminator If the fault persists the unit needs servicing 6 2 1 8 LAMP FAULT Lamp will not strike or stay lit Replace the lamp If the fault persists the unit needs servicing...

Страница 14: ...screen will display The fan will run to reduce the temperature If acceptable reductions achieved within ten seconds the lamp will turn back on to resume normal operation 6 2 1 12 COOLING SHUTDOWN If c...

Страница 15: ...screen Lamp Hours 950 Intensity Adjust 100 _ MENU STANDBY NOTE The brightness control operates a progressive electro mechanical shutter The lamp is running at full power if the power switch is on no m...

Страница 16: ...0 hours 6 2 2 5 CHANGE FILTER AND RESET Lamp Hours 950 Change Filter and Reset Intensity Adjust 100 _ MENU STANDBY The Change Filter and Reset message is displayed flashing whenever the unit runtime h...

Страница 17: ...o the left of CHANGE FILTER AND RESET After installing a new filter touch RESET FILTER to turn off the flashing signal for the next 250 hours of unit operation 6 2 2 6 2 ERROR LOG E1 _ Error Log MENU...

Страница 18: ...d PIC The minus or plus controls will cycle between them 6 2 2 6 5 LANGUAGE English MENU Language 20 _ Lamp Hours The minus or plus controls will cycle through the choices of variation languages Press...

Страница 19: ...enings 8 MAINTENANCE SERVICING AND REPAIR WARRANTY Performance of routine maintenance is essential to proper operation of the illuminator Follow the steps below for replacing filters and lamps Regular...

Страница 20: ...rip flanges of the plastic housing of the lamp pull straight out to remove the lamp Insert a new lamp pushing firmly to assure full engagement to the power connectors Lower the lamp engagement lever c...

Страница 21: ...to a new owner without authorization from NICO Corporation This warranty gives you specific legal rights and you may also have other rights that vary from state to state POST WARRANTY REPAIRS Please...

Страница 22: ...the European Community Caution Caution Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner Consult Instructions for Use CE mark Not disposable in general...

Отзывы: