16
3.
PANELS AND I/O TERMINALS
MULTIFUNCTION GENERATOR
3.2.6 Frequency reference output (REF OUT)
Use REF OUT to unify the frequency and phase for multiple WF1973, WF1974 units.
Connect the frequency reference output of the master unit when multiple units are
combined in synchronous connection, or the master WF1973 or WF1974, to the external 10
MHz frequency reference input of the WF1973 or WF1974.
)
“4. SYNCHRONIZING MULTIPLE UNITS” in the Application Instruction Manual
■
Output Characteristics
3.2.7 Multi-I/O (MULTI I/O)
MULTI I/O can be used for sweep control and sequence control. It outputs the step
synchronization code for the sequence.
Input frequency
10 MHz (±0.5% (±50 kHz))
Input waveform
Sine or square (50 ±5% duty)
Signal GND
Insulated from housing and each channel waveform output (42 Vpk max.)
Output voltage
1 Vp-p / 50
Ω
Output impedance
50
Ω
, AC coupled
Output frequency
10 MHz
Output waveform
Square
Signal GND
Same potential as housing
WF1974
WF1973
REF OUT
REF OUT
Check
Do not connect any equipment other than the WF1973 and WF1974 not
specified by NF Corporation to the frequency reference output.
The special signal that is output from this terminal during synchronization may
make the operation of such connected equipment unstable.
WF1974
WF1973
MULTI I/O
MULTI I/O
Содержание WF1973
Страница 1: ...MULTIFUNCTION GENERATOR WF1973 WF1974 Instruction Manual Basics NF Corporation...
Страница 2: ...MULTIFUNCTION GENERATOR WF1973 WF1974 Instruction Manual Basics DA00016809 001...
Страница 169: ...http www nfcorp co jp 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN NF Corporation...