5.3
Explanation of Commands
DF1906
5-14
・
:SYSTem :PON[:OUTPut] {POFF|PON|LAST}
:SYSTem :PON[:OUTPut]?
Explanation: Sets/queries the output state immediately after power on.
Parameter:
POFF (Power off)
PON (Power on)
LAST (Output state at power off)
Response: Returns any of POFF, PON, and LAST.
ex)
SYST:PON?
POFF
・
:OUTPut:IMPe dance {0 |1}
:OUTPut:IMPeda nce?
Explanation: Sets/queries the load impedance for converting the amplitude/DC offset
(Hi/Lo level) display value.
Parameter:
0 (open)
1 (50 Ω)
Response: Returns 0 or 1.
ex)
OUTP:IMP?
1
・
:SYSTem :ERRor?
Explanation: Queries the error state.
Parameter: None
Response: Returns the error number. Returns 0 if no error occurred.
Error Number
“