3.5.5 Set Amplitude and DC Offset
"Amplitude" indicates the AC amplitude of an output signal, and "DC offset" indicates the median of an
output signal.
When H ( High Level) and L (Low Level) are displayed,
"
4.3.4 Set by High Level/Low Level
"
For load setting,
"
4.3.3 Select Load
"
When you press amplitude/offset setting key [
AMP/OF S]
, the cursor moves to the amplitude setting
value (on the right of the "
A
" display) on the display panel. When you press the key once again, the
cursor moves to the DC offset setting value (on the r ight of the "O" display) on the display panel. (In
the DC mode, the amplitude is zer o and the cursor moves to the DC offset setting value.)
Amplitude setting range: 0 to 10 Vp-p/open (0 to 5 Vp-p/50
Ω), resolution: 1 mV. (Resolution is 1 mV,
regardless of the Vp-p/Vrms setting and load setting.)
DC offset setting range: ±5 V/open (±2.5 V
/50 Ω), resolution: 1 mV. (Resolution is 1 mV, regardless of
the load setting.)
When you turn the [
MODIF Y
] dial when the cursor is located in the amplitude setting value pos ition,
the amplitude can be set. When you turn the [
MODIF Y]
dial when the cursor is located in the DC offset
setting value pos ition, the DC offset can be set.
You can move the cursor position left and r ight us ing cursor movement keys ([
▲
]/[
▼
] keys).
When you operate the [
MODIFY
] dial, any of the follow ing oper ations is performed depending on the
cursor position.
・
When the cursor is located in the value pos ition, the value in that digit pos ition increases/decreases.
・
When the cursor is located in the amplitude unit position ( Vp-p/Vrms display) (only when the
waveform is a sine wave),
the units are toggled between Vp-p (peak-peak voltage) and Vrms (effective voltage).
◎
Example of an operation to set the amplitude to 2 Vrms
・
Press the [
AMP/OFS
] key to enter the amplitude setting mode.
・
Press the [
▼
] key to move the cursor to Vp-p.
・
Turn the [
MODIF Y
] dial to switch to Vrms.
・
Press the
[
▲
] key to move the cursor to 1Vrms.
・
If the digit after 1Vrms is not zero, press and release the [
TRIG / CLEAR
] key quickly while
holding down the [
▼
] key
to clear the value. Then, press the [
▲
] key to move the cursor back to 1Vrms.
・
Turn the [
MODIF Y
] dial to change "
1
" in 1Vrms to "
2
".
If the maximum s ignal output with open load exceeds ±5 V as a result of the amplitude/DC offset
setting, the output may be clipped. The maximum output can be expressed as ±
{(
amplitude [ Vp-p]÷2
)
+
|DC offset [V]|
}
. (Except for when the positive and negative maximum waveform data value is not
used in arbitrary waveforms)
F
500.0000000kHz
A
8.000Vp-p
O
-1.000V
<CONT>
Duty
50.0%
/OPEN
OFF
When cursor is locat ed here,
set the amplitude
When cursor is locat ed here,
set the DC offs et
When waveform is sine wave and curs or is located here,
select the amplitude unit (Vp-p/Vrms)
Select load
Содержание DF 1906
Страница 1: ...NF Corporation DIGITAL FUNCTION GENERATOR DF1906 INSTRUCTION MANUAL ...
Страница 2: ......
Страница 3: ...DIGITAL FUNCTION GENERATOR DF1906 INSTRUCTION MANUAL DA00057652 001 ...
Страница 8: ...DF1906 iv ...
Страница 13: ...DF1906 1 1 1 Overview 1 1 Features 1 2 1 2 Applications 1 2 1 3 Function List 1 3 1 4 Operating Principle 1 4 ...
Страница 26: ...Rear Panel DF1906 3 4 Blank ...
Страница 28: ...Rear Panel DF1906 3 6 Blank ...
Страница 42: ...DF1906 3 20 Blank ...
Страница 70: ...DF1906 5 20 Blank ...
Страница 74: ...DF1906 6 4 ...
Страница 90: ......
Страница 92: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508JAPAN ...