3.5
Basic Operations
DF1906
3-16
3.5.3 Select Setting Value
■ Cursor
The cursor is also displayed in the waveform and oscillation mode. However, the cursor is invalid.
■ Modify
When you turn the [
MODIFY]
dial, the option in the item where the cursor is displayed changes.
3.5.4 Set Frequency
When you press the frequency setting key [
FREQUENCY]
, the cursor moves to the frequency setting
value (on the right of the "
F
" display) on the display panel.
When you turn the [
MODIFY]
dial, the frequency can be set.
You can move the cursor position left and r ight us ing cursor movement keys ([
▲
]/[
▼
] keys).
When you operate the [
MODIF Y
] dial, any of the follow ing operations is performed depending on
the cursor position.
・
When the cursor is located in the value pos ition, the value in that digit pos ition
increases/decreases.
・
When the cursor is located in the subunit position (on the immediate left of "
Hz
"), the subunit is
switched to another one.
◎
Example of an operation to set the frequency to 500 kHz
・
Press the [
FREQUENCY
] key to enter the frequency setting mode.
・
Press the [
▲
] key to move the cursor to 100 kHz.
・
If the digit after 10 kHz is not zero, press and release the [
TRIG / CLEAR
] key quickly while
holding down the [
▼
] key to clear the value. Then, press the [
▲
] key to move the cursor back
to 100 kHz.
・
Turn the "
MODIF Y
" dial to change "
1
" in 100 kHz to "
5
."
◎
Example of an operation to set the frequency display to unit kHz
・
Press the [
FREQUENCY
] key to enter the frequency setting mode.
・
Press the [
▼
] key to move the cursor to the position immediately left of Hz on the right of the
value part.
・
Turn the [
MODIF Y
] dial to switch between Hz / kHz / MHz / mHz.
Frequency setting range: 0.1 mHz to 2 MHz, resolution: 0.1 mHz.
The rated frequency for a ramp wave, square wave w ith a duty rate of other than 50%, arbitrary
waveform, and burst/trigger/gate oscillation mode is up to 50 kHz, but up to 2 MHz can be set.
F
500.0000000kHz
A
8.000Vp-p
O
-1.000V
<CONT>
Duty
50.0%
/OPEN
OFF
When cursor is located her e,
select the subunit
When cursor is located her e,
set the frequency
Содержание DF 1906
Страница 1: ...NF Corporation DIGITAL FUNCTION GENERATOR DF1906 INSTRUCTION MANUAL ...
Страница 2: ......
Страница 3: ...DIGITAL FUNCTION GENERATOR DF1906 INSTRUCTION MANUAL DA00057652 001 ...
Страница 8: ...DF1906 iv ...
Страница 13: ...DF1906 1 1 1 Overview 1 1 Features 1 2 1 2 Applications 1 2 1 3 Function List 1 3 1 4 Operating Principle 1 4 ...
Страница 26: ...Rear Panel DF1906 3 4 Blank ...
Страница 28: ...Rear Panel DF1906 3 6 Blank ...
Страница 42: ...DF1906 3 20 Blank ...
Страница 70: ...DF1906 5 20 Blank ...
Страница 74: ...DF1906 6 4 ...
Страница 90: ......
Страница 92: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508JAPAN ...