5.3
Explanation of Commands
DF1906
5-12
・
[:SOURce]:BURSt: SPACes <num space>
[:SOURce]:BURSt:SPACes?
Explanation: Sets/queries the stop wave number for burst oscillation.
Parameter: Stop wave number
0.5 to 256
Response: Returns the set stop wave number.
Numeric data is transferred as ASCII byte in NR2 format.
ex)
BURS:SPAC?
10.0
Remark: The stop wave number can be set in increments of 0.5 wave for burst oscillation
when both the mark wave number and space wave number are 128 or less.
・
[:SOURce]:PHASe <phase>
[:SOURce]:PHASe?
Explanation: Sets/queries the oscillation stopping phase for burst oscillation, etc.
Parameter: Phase [deg]
-180.0(-180deg) to +180.0(+180deg)
Response: Returns the set phase in [deg].
Numeric data is transferred as ASCII byte in NR2 format.
ex)
PHAS?
90.0
・
:OUTPut:SYNC:TYPE {PHASe |STATe}
:OUTPut:SYNC:TYPE?
Explanation: Sets/queries the
SYNC OUT
output type.
Parameter:
PHASe (Waveform synchronous output)
STATe (Oscillation state synchronous output)
Response: Returns PHAS or STAT.
ex)
OUTP:SYNC:TYPE?
PHAS
・
:TRIGger[:SEQuence][:IMMediate] {0|1}
Explanation: Controls the trigger or gate signal.
Parameter:
In trigger mode
0 (No change)
1 (Trigger input)
In gate mode
0 (Gate off)
1 (Gate on)
Содержание DF 1906
Страница 1: ...NF Corporation DIGITAL FUNCTION GENERATOR DF1906 INSTRUCTION MANUAL ...
Страница 2: ......
Страница 3: ...DIGITAL FUNCTION GENERATOR DF1906 INSTRUCTION MANUAL DA00057652 001 ...
Страница 8: ...DF1906 iv ...
Страница 13: ...DF1906 1 1 1 Overview 1 1 Features 1 2 1 2 Applications 1 2 1 3 Function List 1 3 1 4 Operating Principle 1 4 ...
Страница 26: ...Rear Panel DF1906 3 4 Blank ...
Страница 28: ...Rear Panel DF1906 3 6 Blank ...
Страница 42: ...DF1906 3 20 Blank ...
Страница 70: ...DF1906 5 20 Blank ...
Страница 74: ...DF1906 6 4 ...
Страница 90: ......
Страница 92: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508JAPAN ...