March 2019
30
VeraSol
STATus?
Description
Requests the status of the LSS-7120.
Response
A decimal number containing bit-encoded information:
If bit 0 is set, the output is on.
Bit
1:
reserved
If bit 2 is set, the head is disconnected (check the cable).
If bit 3 is set, the head has not yet warmed up to its optimum operating temperature.
If bit 4 is set, the head temperature is above its optimum operating range.
Example
Response: 8 (binary 00001000) means the head is not yet warm (bit3 = 1), and the output is off (bit0 = 0).
USERCAL:SELECT 1|0
USERCAL:SELECT?
Description
Select or report user-cal selection (factory or user-performed)
Parameters
1 – User calibration offset
0 – Factory Calibration
Reset Value
0
USERCAL:PERFORM
Description
Offsets the current output intensity to indicate 1.0 sun on the display.
Parameters
None