25
The location of firmware files is under the application installation folder, usually:
C:\Program Files\Ophir Optronics\Ophir Ethernet Application\Firmware\
or:
C:\Program Files (x86)\Ophir Optronics\Ophir Ethernet Application\Firmware\
Note:
It is recommended not to upgrade firmware unless a new version is actually available.
If a firmware upgrade is not actually required, the following message will be displayed; answer “no” to abort the
firmware upgrade:
In the event that a new firmware upgrade is available and necessary, the software will continue with the upgrade
automatically.
Note:
Do not disconnect the communications or power cables while the firmware is being upgraded. Firmware
upgrade takes ~5 minutes.
In the event that the firmware upgrade is disrupted before it is finished, the device will start up in a special “Boot”
mode after the next power up. In that case, the firmware upgrade can be repeated and completed using the same
menu and method as above.
Note:
It is recommended that the device be powered off and on again after performing a firmware upgrade, this is
especially important when using DHCP mode.
Zeroing:
Choose the Zero option in the menu:
The following window will appear: