15
Returning to Factory Default Settings:
If connection has been lost to the device, and there is no way to find out the present IP address setting, as a last
resort the user can return the device temporarily to factory default settings in order to establish a connection and
then update the configuration as required.
In order to do this, the PCB needs to be removed from the mechanical box. The simplest way to do this is first to
remove all connectors from the device; then unscrew the two screws on the left and right of the panel where the
D15 “sensor” socket is located; and then pull the PCB out along with the D15 panel.
Once removed, press down the button labeled “SW1”, then connect the power supply and Ethernet connector while
holding the button down - the device will power up with default factory settings. The IP address will be set to default
(see above), which is suitable for connection only via a direct connection to the PC network port. Establish
connection to the device and then change the settings as described above. Then close the box by returning the
screws removed earlier.