7
2.1.3 Installing Table Top
If your table is equipped with the Doubler Interface be sure to follow the
instructions in Section 2.4 for assembling the table sections prior to installing
them on the isolators.
Warning
Table tops are heavy! As described in Section 2.1.1, use a forklift
or other appropriate equipment. Be sure to use proper lifting
procedures to avoid severe personal injury.
A. With the isolators located as described in Section 2.1.2, raise the table
top and position it above the isolators. Be sure each isolator is
centered within the tapped three-hole patterns on the lower surface
of the table (Figure 2.2).
Isolator Clamp
Hex Head bolt
Flat Washer
(3 per Isolator)
Table Bottom
Surface
Figure 2.2 — Isolator Clamps and Mounting Holes
B. Gently lower the table so that is just above the top of isolators.
C. Install three isolator clamps in the corners where each isolator is
located (Figures 1.1 & 2.2). Loosely secure the clamps with the bolts.
Do not tighten the clamps at this time.
D. Lower the table top until it rests on the isolator support plates.
Caution
When lowering the table on to the supports, do not allow the table
to shift sideways. Doing so could damage the isolators.
Содержание IPV
Страница 2: ...ii...
Страница 22: ...16 Figure 13 Double Bolt Tightening Sequence...
Страница 32: ...INSTRUCTION MANUAL Vibration Control System...
Страница 39: ...viii...
Страница 63: ...24...
Страница 64: ...25...