5
2 Assembly
2.1
Assembling the System
2.1.1
Table Installation Assistance
Optical tables are heavy. Even a small table top weighs over 500 pounds
(227 Kg). Contact Newport or your local sales representative for help in
preparing your facility.
2.1.2
Positioning the Isolators
Figure 1 shows the correct isolator positions for standard table sizes. Table 1
shows the correct dimensions (A, B, & C) for each table size.
Figure 1
— Isolator Position
Table 1
— Isolator Positions
Table
Width
ft
(mm)
Table
Length
ft
(mm)
A
in
(mm)
B
in
(mm)
C
in
(mm)
3
(900)
6
(1800)
14
(341)
30
(733)
35.5
(899)
3
(900)
8
(2400)
14
(341)
42
(1028)
47.25
(1159)
3
(900)
10
(3000)
14
(341)
56
(1374)
60.25
(1479)
4
(1200)
6
(1800)
14
(341)
30
(733)
43.75
(1079)
4
(1200)
8
(2400)
26
(641)
42
(1028)
52.25
(1310)
4
(1200)
10
(3000)
26
(641)
56
(1374)
65.25
(1604)
4
(1200)
12
(3600)
26
(641)
70
(1720)
77.75
(1914)
4
(1200)
14
(4200)
26
(641)
84
(2166)
90.75
(2326)
4
(1200)
16
(4800)
26
(641)
94
(2311)
100
(2463)
5
(1500)
6
(1800)
38
(941)
30
(733)
51.75
(1296)
5
(1500)
8
(2400)
38
(941)
42
(1028)
60
(1499)
5
(1500)
10
(3000)
38
(941)
56
(1374)
71
(1766)
5
(1500)
12
(3600)
38
(941)
70
(1720)
82.75
(2054)
5
(1500)
14
(4200)
38
(941)
84
(2166)
95
(2445)
5
(1500)
16
(4800)
38
(941)
94
(2311)
104.25
(2576)
Содержание IPV
Страница 2: ...ii...
Страница 22: ...16 Figure 13 Double Bolt Tightening Sequence...
Страница 32: ...INSTRUCTION MANUAL Vibration Control System...
Страница 39: ...viii...
Страница 63: ...24...
Страница 64: ...25...