6.1
Installation Location & Ventilation
The FSM-CD300B controller/driver is designed for indoor operation
in an ambient temperature of 0 to 35°C (32 to 95°F). Component cooling
is provided by a fan, which aspirates air through slots in both sides of the
unit and ejects air through the back. To assure adequate airflow, provide a
minimum clearance of 25 mm (1”) on both sides of the unit and 2” (50 mm)
in back of the unit. Also, adequate spacing behind the fan provides quieter
operation.
6.2 Electrical
Connections
The FSM mirror head is interfaced to the FSM-CD300B controller/driver by
the system’s 15-pin connector cable. The controller/driver is powered from
an AC wall outlet. It is equipped with a universal power supply that
accommodates 100-240 Vac, 50/60 Hz. No switch or fuse needs to be
changed when going from 100 to 120 or 240 Vac power.
Prior to applying AC power, verify that the protective cover and protective
packing material have been removed from the FSM mirror head.
Prior to applying AC power, connect the system’s15-pin connector cable to
the FSM mirror head and FSM-CD300B controller/driver. If you later need to
remove the 15-pin connector cable, first remove AC power. Connecting and
disconnecting the 15-pin connector cable in the absence of power will avoid
making or breaking powered signal connections.