MCS260B
CORNERSTONE 260B MONOCHROMATORS
64
403
Cannot Reach
Wavelength
Typically, when grating-change command
results in unobtainable wavelength.
404
Grating Halted
Caused by grating or wavelength command
during firmware upgrade. Also, by rotation stage
problem.
501
Filter Wheel Missing
Filter wheel not detected at power-up.
502
Filter Wheel Fault
Hardware failure: filter wheel not operating
properly.
504
Filter Halted
Filter command during firmware upgrade.
505
Slit Controller Missing
Slit command received but motorized slit not
connected.
506
Slit Controller Fault
A problem was detected with a motorized slit.
507
Slit Controller Halted
Caused by slit command during firmware
upgrade.
508
Unable To Home Slit
Could not find home on a motorized slit.
Typically performed at power-up.
510
Unable To Home Turret
Could not find home on grating turret.
512
Turret Halted
Grating command during firmware upgrade. Also
caused by turret rotation problem.
520
Accessory Missing
Command received for uninstalled accessory;
e.g. “outport” command with no flip mirror.
531
Mirror End Not Detected
Flip mirror motor fault.