VENTILATOR SET UP AND USE
5-12
OPRHT50NA B0908
Set Up And Operation
1. From the Off/Standby condition, switch the ventilator to the
Setting condition by pressing On/Standby one time.
2. Turn the humidifier on by pressing the Humidifier On button.
3. Adjust humidification temperature.
4. After a minimum of 30 minutes (for temperature to stabilize),
maximum of 55 minutes, start ventilation by pressing
On/Standby again.
WARNING
Failure to change from the Setting condition to
the On condition within 60 minutes will cause the humidifier to
turn off. After restarting, ensure that temperature is
appropriate before starting ventilation.
5. When the humidifier water level reaches the “REFILL” line,
refill with sterile distilled water. The humidifier must be
restarted afterward by pressing the Humidifier On button.
6. The absorbent paper in the humidifier should be replaced
with each cleaning, weekly during use, or sooner if it
becomes torn.
NOTE:
Humidifier specifications may be found on pg 2-7.
NOTE:
For cleaning, disinfection and sterilization of the
humidifier, please see Section 6.
OXYGEN ACCESSORIES
WARNING
Continuous oxygen monitoring is required for
patient safety. The HT50 does not have a built-in alarm
system to notify user of a failure or disconnection of the
oxygen source.
WARNING
Ensure that the oxygen source is not empty
before and during the use of Air/Oxygen Entrainment Mixer or
Oxygen Blending Bag Kit.
Содержание Newport HT50
Страница 26: ...SECTION 3 OPRHT50NA B0908 3 3 Figure 3 1 HT50 Ventilator Front Panel model HT50 H1 HT50 H1B ...
Страница 78: ...OPRHT50NA B0908 6 CLEANING MAINTENANCE Sterilization Cleaning 6 1 Maintenance 6 9 General Warnings 6 11 ...
Страница 112: ...OPRHT50NA B0908 APPENDIX B HT50 ACCESSORIES HT50 Ventilator Parts and Accessories B 1 ...
Страница 115: ...OPRHT50NA B0908 APPENDIX C HT50 HELP GUIDE Questions and Answers C 1 ...