SECTION 7
OPRHT50NA B0908
7-13
Mixer makes a pronounced
clicking sound during
normal operation
Oxygen source gas pressure is low.
Oxygen source regulator is oscillating.
Check that oxygen source gas is 50 psig.
Check oxygen source regulator. If noise
continues contact Newport Technical
Service Dept.
Oxygen leaks out of
Mixer when connected
to 50 psig oxygen gas
source.
Mixer diaphragm is leaking.
Contact Newport Technical Service Dept.
TABLE 7-5: AIR OXYGEN ENTRAINMENT MIXER
PROBLEM
POTENTIAL CAUSE
SUGGESTED CORRECTIVE ACTION
Monitored FIO2 is lower
than set FIO2 by > .08
when using Air Oxygen
Entrainment Mixer
Filter cover is loose.
Filter cover needs to be replaced.
Tighten filter cover.
Contact your provider or Newport to obtain
replacement filter cover.