SECTION 2
OPRHT50NA B0908
2-7
User Set Up:
Power Save
On
Pressure Units
cmH
2
O
Auto Panel Lock
Disabled
Set Up
User
42. Patient Circuit
Reusable 22 mm I.D. adult/pediatric circuit with 3/16 inch
(4.8 mm) I.D. proximal pressure sensing line, 1/8 inch (3.2 mm)
I.D. exhalation valve control drive line, and exhalation valve.
43. Exhalation Valve
N
EWPORT
'
S
HT50 exhalation valve (P/N HT600039) is
manufactured and designed specifically for the N
EWPORT
HT50
Ventilator. Newport Medical does not approve of the use of
any type or brand of exhalation valve that has not been tested
and approved by Newport Medical for use with the HT50.
HT50-H, HT50-HB Humidifier Specifications
(operates on A.C. power only)
Set Target Temperature Range:
19ºC to 39ºC
Operating Water Volume:
300 mL
Usable Volume of
Water Bottle:
265 mL
Compliance at Minimum
Water Level (Refill Line):
0.5 mL/cmH
2
O / mbar @ 23ºC
Compliance at Maximum
Water Level (Full Line):
0.33 mL/cmH
2
O / mbar @ 23ºC
Intended Use:
Adult and pediatric patients whose supraglottic airway is or is
not bypassed.
Warm-Up Time:
30 minutes
Gas Leakage:
2 mL/min at airway pressure of 80 cmH
2
O / mbar
Humidifier Output:
33.8 mg/L at a continuous flow of 10 L/min @ 39ºC
Maximum Operating
Airway Pressure:
100 cmH
2
O / 98 mbar
Maximum Temperature at
the Patient Wye That
Triggers an Alarm:
41ºC
Содержание Newport HT50
Страница 26: ...SECTION 3 OPRHT50NA B0908 3 3 Figure 3 1 HT50 Ventilator Front Panel model HT50 H1 HT50 H1B ...
Страница 78: ...OPRHT50NA B0908 6 CLEANING MAINTENANCE Sterilization Cleaning 6 1 Maintenance 6 9 General Warnings 6 11 ...
Страница 112: ...OPRHT50NA B0908 APPENDIX B HT50 ACCESSORIES HT50 Ventilator Parts and Accessories B 1 ...
Страница 115: ...OPRHT50NA B0908 APPENDIX C HT50 HELP GUIDE Questions and Answers C 1 ...