SECTION 5
OPRHT50NA B0908
5-3
e. For use without a humidifier or HME, attach the other end
of the Proximal Pressure line to the pressure port on the
exhalation valve. See Detail A, Figure 5–1.
• For use with an HME, attach the Proximal Pressure line
to a pressure tee adapter (p/n NP130-22) on the patient
side of the HME. Use the cap that comes with the tee
adapter to seal the pressure port on the exhalation
valve. See Figure 5–5.
• For use with a humidifier, attach the Proximal Pressure
line to a pressure tee adapter (p/n NP130-22) before
the inlet port on the humidifier. Use the cap that
comes with the tee adapter to seal the pressure port
on the exhalation valve. See Figure 5–4.
f. Attach the reusable exhalation valve to the patient end of
the breathing circuit.
g. Attach one end of the exhalation valve drive line to the
exhalation valve and the other end to the Exh. Valve
connection located below the Gas Output on the HT50.
See Figure 5–1.
h. When using the HT50 built-in humidifier (model HT50-H,
HT50-HB), plug the temperature probe electrical connector
into the Temp Probe connection on the left side of the
HT50. Attach the short wire plug into the elbow outlet on
top of the humidifier. Attach the long wire plug into the
exhalation valve port. See Figure 5–2.
Exhalation Valve Calibration
6.
Calibrate the exhalation valve prior to use as follows:
WARNING
Each time an exhalation valve is replaced or
maintained it must be calibrated.
a. Connect the HT50 patient breathing circuit to the
ventilator as instructed above. Connect an adult (500 mL)
test lung with restrictor (LNG500A) or occlude the patient
connection of the breathing circuit.
b. Press the On/Standby button once to enter Settings
condition.
c. Press the Manual Inflation button once, then again within
three seconds.
d. The HT50 will start the EZ Cal and the ventilator will
automatically test the exhalation valve. If it passes the
test, the messages "Cal Completed", then "Press ON to
Vent" will be displayed.
Содержание Newport HT50
Страница 26: ...SECTION 3 OPRHT50NA B0908 3 3 Figure 3 1 HT50 Ventilator Front Panel model HT50 H1 HT50 H1B ...
Страница 78: ...OPRHT50NA B0908 6 CLEANING MAINTENANCE Sterilization Cleaning 6 1 Maintenance 6 9 General Warnings 6 11 ...
Страница 112: ...OPRHT50NA B0908 APPENDIX B HT50 ACCESSORIES HT50 Ventilator Parts and Accessories B 1 ...
Страница 115: ...OPRHT50NA B0908 APPENDIX C HT50 HELP GUIDE Questions and Answers C 1 ...