OPERATOR’S RESPONSIBILITY
1-4
OPRHT50NA B0908
Section 2). Observe all of the
WARNINGS
and
Cautions
posted in
this manual and on labels found on the HT50 Ventilator and
associated accessories.
General Warnings
External power connection: To maintain grounding integrity when
using A.C. power, only connect to hospital grade receptacles.
Always disconnect the external power supply prior to servicing.
Always use the A.C. power cord supplied with the HT50. Make
certain the power cord ferrite is always attached to the A.C. power
cord to ensure that the HT50 meets EMC requirements.
There is a risk of explosion if used in the presence of flammable
anesthetics.
All settings and adjustments in the different ventilation modes must
be made in accordance with a physician’s prescribed therapy.
N
EWPORT
cannot warrant or endorse the safe performance of third
party humidifiers for use with the HT50.
When the HT50 is operating on battery power, the optional built-in
humidifier does not function. A heat moisture exchanger, or other
humidification device, should be used until the unit is connected to
A.C. power at which time the built-in humidifier can be used.
Do not use electrically conductive patient circuits.
Always use a clean patient circuit.
Always use an inline filter (p/n HT6004701 or equivalent) at the
Airway Pressure Connector to protect the internal transducers from
moisture or other contaminants.
Always use appropriate monitors to ensure sufficient oxygenation
and ventilation (such as pulse oximeter and/or capnograph) when
the HT50 Ventilator is in use on a patient.
The ventilator is ready for operation only when:
a) It is completely assembled, and;
b) The Quick Check Procedure, including the Exhalation Valve
Calibration (see Appendix A, Operating Manual) or OVP
(Service Manual) has been successfully completed.
Constant attention by qualified medical personnel is recommended
whenever a patient is ventilated with the HT50.
When the HT50 is used in homecare environments, proper
education and training of the appointed caregiver must be
provided prior to the patient leaving the health care facility.
If a fault is detected in the ventilator and its life support functions
are in doubt, immediately discontinue use; use an alternative
Содержание Newport HT50
Страница 26: ...SECTION 3 OPRHT50NA B0908 3 3 Figure 3 1 HT50 Ventilator Front Panel model HT50 H1 HT50 H1B ...
Страница 78: ...OPRHT50NA B0908 6 CLEANING MAINTENANCE Sterilization Cleaning 6 1 Maintenance 6 9 General Warnings 6 11 ...
Страница 112: ...OPRHT50NA B0908 APPENDIX B HT50 ACCESSORIES HT50 Ventilator Parts and Accessories B 1 ...
Страница 115: ...OPRHT50NA B0908 APPENDIX C HT50 HELP GUIDE Questions and Answers C 1 ...