SECTION 3
OPRHT50NA B0908
3-9
If the Volume Control setting causes the flow rate to reach the
maximum or minimum level of the flow specification, adjustment
of Volume Control ceases, a beep sounds, and a setting
limitation message appears in the Message Display Window.
NOTE:
See pg. 3-24 for a list of setting limitation messages.
NOTE:
In SPONT mode, the Volume Control is not utilized but
the value can be preset.
Switching from Pressure Control to Volume Control:
Press the Volume Control button. The set tidal volume is displayed
in the adjacent window if the HT50 is ventilating. A “PRESS
AGAIN” message appears in the Message Display Window.
Adjust the tidal volume level by pressing the
▲
Up/
▼
Down controls
while the LED and numerical display are blinking. If the HT50 is
ventilating you will need to press the Volume Control button again
within 5 seconds following adjustment.
Pressure Control
(target pressure)
Range:
5 to 60 cmH
2
O / mbar
NOTE:
When Pressure Control is first initiated or the setting is
changed, the first few breaths may cycle off early until slope/rise
is optimized. If early cycling off continues, re evaluate the
breathing circuit configuration and lengthen the tubing as
necessary.
Pressing this control button, followed by pressing the
▲
Up/
▼
Down controls, allows the adjustment of the target airway
pressure setting. Target pressure is referenced to ambient
(atmospheric pressure). When the green Pressure Control LED
illuminates, the adjacent window displays the set airway
pressure. See Theory of Operation pg 4-4 for more details.
NOTE:
In SPONT mode, the Pressure Control is not utilized but
the value can be preset.
Switching from Volume Control to Pressure Control:
Press the Pressure Control button. The set target airway pressure
value is displayed in the adjacent window if the HT50 is
ventilating. A “PRESS AGAIN” message appears in the Message
Display Window.
Adjust the set target airway pressure by pressing the
▲
Up/
▼
Down controls while the LED and numerical display are
blinking. If the HT50 is ventilating you will need to press the
Pressure Control button again within 5 seconds following
adjustment.
Содержание Newport HT50
Страница 26: ...SECTION 3 OPRHT50NA B0908 3 3 Figure 3 1 HT50 Ventilator Front Panel model HT50 H1 HT50 H1B ...
Страница 78: ...OPRHT50NA B0908 6 CLEANING MAINTENANCE Sterilization Cleaning 6 1 Maintenance 6 9 General Warnings 6 11 ...
Страница 112: ...OPRHT50NA B0908 APPENDIX B HT50 ACCESSORIES HT50 Ventilator Parts and Accessories B 1 ...
Страница 115: ...OPRHT50NA B0908 APPENDIX C HT50 HELP GUIDE Questions and Answers C 1 ...