Finally apply the calibration gas and check to
see if the calibration concentration is indicated
correctly
If the indicated value is outside the auto zero adjustment range when the RESET key is pressed, the message
window will display a blinking “Err” indication and auto zero adjustment will be impossible. If this occurs, carry
out analog span adjustment in maintenance mode 2.
MEMO
Note that a gas leak alarm and external output will be generated when calibration gas is applied in normal
mode.
CAUTION
POWER
ALARM
2
1
RESET
BZ STOP
Check the
calibration
concentration.