- 21 -
2) Analog span adjustment
(
Maintenance mode 2
)
①
Set the Mode switch at 2 to select maintenance
mode 2.
②
Press down the RESET key until the Power
lamp goes off.
③
The Power lamp lights up again and the auto
span adjustment value is reset.
④
Set the Mode switch at 1 to select maintenance
mode 1.
⑤
Confirm that the zero point is at 1.0 and apply
calibration gas to the gas detector for a minute.
⑥
Set the Mode switch at 2 to select maintenance
mode 2.
Always perform analog zero adjustment before doing analog span adjustment. Otherwise the span cannot be
adjusted properly when the analog span adjustment is performed.
CAUTION