TABLE 1
SETTINGS OF SWITCH 1 (S1) ON THE CONTROL BOARD
1.
S1-2 and S1-1 are used for setting baud rate:
Baud
Rate
S1-2
S1-1
1200
ON
ON
2400
ON
OFF
4800
OFF
ON
9600
OFF
OFF
2.
S1-3 is used for setting parity check:
Parity
S1-3
even
ON
no
OFF
3.
S1-4 is used for setting mode:
Mode
S1-4
multidrop
ON
non-multidrop
OFF
4.
S1-8, S1-7, S1-6, S1-5 are used for setting unit number:
Unit
#
S1-8
S1-7
S1-6
S1-5
0
ON
ON
ON
ON
1
ON
ON
ON
OFF
2
ON
ON
OFF
ON
3
ON
ON
OFF
OFF
4
ON
OFF
ON
ON
5
ON
OFF
ON
OFF
6
ON
OFF
OFF
ON
7
ON
OFF
OFF
OFF
8
OFF
ON
ON
ON
9
OFF
ON
ON
OFF
10
OFF
ON
OFF
ON
11
OFF
ON
OFF
OFF
12
OFF
OFF
ON
ON
13
OFF
OFF
ON
OFF
14
OFF
OFF
OFF
ON
15
OFF
OFF
OFF
OFF
M1226-0050 28
Содержание BIOFLO III
Страница 2: ......
Страница 7: ......
Страница 9: ...M1226 0050 2...
Страница 37: ...M1226 0050 30...
Страница 40: ...M1226 0050 33...
Страница 48: ...FIGURE 1 VESSEL SCHEMATIC M1226 0050 41...
Страница 49: ...FIGURE 2 SAMPLING SYSTEM M1226 0050 42...
Страница 50: ...FIGURE 3 FLOWMETER CONVERSION M1226 0050 43...
Страница 51: ...FIGURE 4 FRONT PANEL M1226 0050 44...
Страница 52: ...FIGURE 5 REAR OF CABINET M1226 0050 45...
Страница 53: ...FIGURE 6 SWITCH PANEL M1226 0050 46...
Страница 54: ...FIGURE 7 EXTERNAL NUTRIENT PUMP TERMINAL M1226 0050 47...
Страница 55: ...FIGURE 8 SUGGESTED IMPELLER LOCATION M1226 0050 48...
Страница 56: ...FIGURE 9 HEADPLATE M1226 0050 49...
Страница 57: ...FIGURE 10 SPARGER TUBE M1226 0050 50...
Страница 58: ...FIGURE 11 HARVEST TUBE M1226 0050 51...
Страница 59: ...FIGURE 12 THERMOWELL M1226 0050 52...
Страница 60: ...FIGURE 13 FOAM PROBE M1226 0050 53...
Страница 61: ...FIGURE 14 BAFFLE INSTALLATION M1226 0050 54...
Страница 62: ...FIGURE 15 SIDE PORTS M1226 0050 55...
Страница 63: ...FIGURE 16 pH PROBE INSTALLATION M1226 0050 56...
Страница 64: ...FIGURE 17 D O PROBE PHOENIX INSTALLATION M1226 0050 57...
Страница 65: ...FIGURE 18 D O PROBE INGOLD INSTALLATION M1226 0050 58...
Страница 66: ...FIGURE 19 CONVERSION OF D O PORT TO INOCULATION PORT M1226 0050 59...
Страница 67: ...FIGURE 20 EXHAUST CONDENSER INSTALLATION M1226 0050 60...
Страница 68: ...FIGURE 21 RESTRICTOR IN CONDENSER WATER IN LINE M1226 0050 61...
Страница 69: ...FIGURE 22 DISASSEMBLY OF CONDENSER M1226 0050 62...
Страница 70: ...FIGURE 23 FACTOR AFFECTING pH CONTROL M1226 0050 63...
Страница 86: ...M1226 0050 79...
Страница 87: ...FIGURE 25 BEARING HOUSING ASSEMBLY M1226 0050 80...
Страница 88: ...FIGURE 26 BEARING HOUSING IN STERILIZATION M1226 0050 81...
Страница 89: ...M1226 0050 82...
Страница 90: ...M1226 0050 83...
Страница 91: ...M1226 0050 84...
Страница 92: ...M1226 0050 85...