CHAPTER 1
INTRODUCTION
1.1
SCOPE OF MANUAL
This manual contains a description of the vessel, installation, operating and
maintenance instruction for the BioFlo III System 1.25, 2.5 and 5L Models
manufactured by the New Brunswick Scientific Co., Inc., 44 Talmadge Road,
P.O. Box 4005, Edison, New Jersey 08818-4005, USA.
1.2
DESCRIPTION OF EQUIPMENT
BioFlo III is a versatile bioreactor that provides a fully equipped state of the art
fermentation system in one compact package. BioFlo III can be employed for batch
or continuous culture with microprocessor control of pH, DO, agitation, temperature,
nutrient feed, and electronic foam control.
1.3
DESCRIPTION OF VESSEL
The vessel parts consist of a stainless steel head plate, a flanged glass tube (thick
walled) vessel body which is detachable from the bottom dished head. The dished
head is jacketed for circulation of temperature controlled water. Four sterilizable
polypropylene compression ports are provided in the glass wall for the addition of
antifoam and nutrients, as well as for vessel overflow in continuous culture studies.
Ports are provided in the headplate for: Inoculation, base and acid addition, a
thermowell for Resistance Temperature Detector, a Sparger, a Harvest Tube, a
Sampler, an Exhaust Condenser, Dissolved Oxygen and pH Electrodes. The drive
bearing housing is also located on the headplate.
1.4 AGITATION
SYSTEM
A removable agitation motor located on top of the bearing housing is connected to the
agitation shaft with a multi-jaw coupling. It can be easily disconnected while
autoclaving the vessel and replaced after sterilization. The motor will provide
agitation
speed range of 50 to 1000
±
(2 RPM + 0.5% setting).
1.5 TEMPERATURE
CONTROL
The culture temperature may be selected in the range from 4
°
C to 80
°
C (
±
0.1
°
C) and
is controlled by a microprocessor based PI (Proportional and Integral) controller. The
media temperature is sensed by an RTD (Resistance Temperature Detector)
submerged
in the thermowell.
M1226-0050 1
Содержание BIOFLO III
Страница 2: ......
Страница 7: ......
Страница 9: ...M1226 0050 2...
Страница 37: ...M1226 0050 30...
Страница 40: ...M1226 0050 33...
Страница 48: ...FIGURE 1 VESSEL SCHEMATIC M1226 0050 41...
Страница 49: ...FIGURE 2 SAMPLING SYSTEM M1226 0050 42...
Страница 50: ...FIGURE 3 FLOWMETER CONVERSION M1226 0050 43...
Страница 51: ...FIGURE 4 FRONT PANEL M1226 0050 44...
Страница 52: ...FIGURE 5 REAR OF CABINET M1226 0050 45...
Страница 53: ...FIGURE 6 SWITCH PANEL M1226 0050 46...
Страница 54: ...FIGURE 7 EXTERNAL NUTRIENT PUMP TERMINAL M1226 0050 47...
Страница 55: ...FIGURE 8 SUGGESTED IMPELLER LOCATION M1226 0050 48...
Страница 56: ...FIGURE 9 HEADPLATE M1226 0050 49...
Страница 57: ...FIGURE 10 SPARGER TUBE M1226 0050 50...
Страница 58: ...FIGURE 11 HARVEST TUBE M1226 0050 51...
Страница 59: ...FIGURE 12 THERMOWELL M1226 0050 52...
Страница 60: ...FIGURE 13 FOAM PROBE M1226 0050 53...
Страница 61: ...FIGURE 14 BAFFLE INSTALLATION M1226 0050 54...
Страница 62: ...FIGURE 15 SIDE PORTS M1226 0050 55...
Страница 63: ...FIGURE 16 pH PROBE INSTALLATION M1226 0050 56...
Страница 64: ...FIGURE 17 D O PROBE PHOENIX INSTALLATION M1226 0050 57...
Страница 65: ...FIGURE 18 D O PROBE INGOLD INSTALLATION M1226 0050 58...
Страница 66: ...FIGURE 19 CONVERSION OF D O PORT TO INOCULATION PORT M1226 0050 59...
Страница 67: ...FIGURE 20 EXHAUST CONDENSER INSTALLATION M1226 0050 60...
Страница 68: ...FIGURE 21 RESTRICTOR IN CONDENSER WATER IN LINE M1226 0050 61...
Страница 69: ...FIGURE 22 DISASSEMBLY OF CONDENSER M1226 0050 62...
Страница 70: ...FIGURE 23 FACTOR AFFECTING pH CONTROL M1226 0050 63...
Страница 86: ...M1226 0050 79...
Страница 87: ...FIGURE 25 BEARING HOUSING ASSEMBLY M1226 0050 80...
Страница 88: ...FIGURE 26 BEARING HOUSING IN STERILIZATION M1226 0050 81...
Страница 89: ...M1226 0050 82...
Страница 90: ...M1226 0050 83...
Страница 91: ...M1226 0050 84...
Страница 92: ...M1226 0050 85...