![NETZSCH LFA 447 Nanoflash Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/netzsch/lfa-447-nanoflash/lfa-447-nanoflash_operating-instructions-manual_1675675082.webp)
Appendix
LFA 447
TM
01/09 J:\LFA447\CHAPTER5.DOC
8
glass plate
19.
Remove the glass plate (move it
outwards).
phillips screws
20.
Loosen the phillips screws.
rear left corner
furnace
21.
Remove the furnace (lift it
upwards). (for reassembly of the
300°C furnace: insert the furnace so
that the furnace cable is placed in
the rear left corner)
A
casing
slot
22.
Insert the casing of the matrix
sample holder so that the slot for
the glass plate shows to the front.