Installation
LFA 447
TM
01/09 J:\LFA447\CHAPTER2.DOC
2
Location Requirements
Select
the optimal component arrangement for the space
you have available.
The location of your instrument must meet the
following requirements:
constant temperature conditions (room temperature) to the
extent possible
no direct sunlight on the instruments
dust-free environment to the extent possible
measuring system
A stable, shock-absorbent table with a working surface of
approx. 2 m x 1 m, is required for the measuring part
(measuring part, computer, printer).
thermostat
The thermostat cooling system requires 0.5 m x 0.5 m
space. The cooling system should be placed beside the
table.
ATTENTION!
Ensure that the measuring part is protected against
external agitation! Agitation during the measurement
affects the sensor and influences the measuring results.
An installation schematic of the measuring system follows
on the next page.