![NETZSCH LFA 447 Nanoflash Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/netzsch/lfa-447-nanoflash/lfa-447-nanoflash_operating-instructions-manual_1675675059.webp)
Operating the Instrument
LFA 447
TM
01/09 J:\LFA447\CHAPTER4.doc
8
Calibration materials
reference samples
recommended max.
temp. [°C]
coating
SRM8421 (electrolytic iron)
700°C
sandblasting, no graphite
above 400°C
SRM1461 (steel)
900°C
sandblasting, no graphite
above 400°C
SRM781 (molybdenum)
2000°C
sandblasting, no graphite
PocoAXM (graphite)
2000°C
graphite
Pyroceram (ceramics)
900°C
graphite
Alumina (ceramics)
2000°C
graphite
The reference samples are also used in laser flash apparatus up to 2000°C. A graphite coating up
to 400°C is possible for all reference samples and therefore useable also for the LFA 447; usually
the maximum temperature of a reference sample is defined in the provided tables of properties
(cp, a …); with exception of ceramic reference samples inert atmosphere is necessary above
300°C.