6.
Steps to import an LFA measurement file into analysis
- Create a new database: >Database>Create new empty
- furthermore:
>Database>Explore> ---new window is opened---
>Database>Import LFA 447 File ---please select correct directory---
c:\ngbwin\ta\nanoflash\files
Double click on the newest file (check the date of files)
After this, you have to confirm the following questions with "yes"; the last measurement
should now be visible in the database window > with a double click on the measurement,
these will be loaded into the analysis
>Measurement>Recalculate diffusivity ---please use Cowan model--- (you can obtain further
information on the models from the help menu)
after calculation, the results are visible; please check the duration time
>Shot>Set range for calculation>10 Halftimes
The use of 10 halftimes is recommended! A comparison of measurements using different
duration times –maybe 2, 10 and 20 halftimes- shows different results, to study the influence
please check this!
Содержание LFA 447 Nanoflash
Страница 2: ...01 09 J LFA447 CHAPTER1 DOC Chapter I General Information LFA 447TM...
Страница 18: ...01 09 J LFA447 CHAPTER2 DOC Chapter II Installation LFA 447TM...
Страница 25: ...01 09 J LFA447 CHAPTER3 DOC Chapter III System Components LFA 447TM...
Страница 50: ...01 09 J LFA447 CHAPTER4 doc Chapter IV Operating the Instrument LFA 447TM...
Страница 70: ...Operating the Instrument LFA 447TM 01 09 J LFA447 CHAPTER4 doc 19...
Страница 73: ...01 09 J LFA447 CHAPTER5 DOC Chapter V Appendix LFA 447TM...