Operating the Instrument
LFA 447
TM
01/09 J:\LFA447\CHAPTER4.doc
13
2.
Check measurement parameters
>Measurement>Parameters
Please use for first measurements 270V, medium and always 300 points for baseline
•
A reason to increase the flash lamp power (>270 V, long pulse) is a bad signal-to-noise
ratio (maybe 8V signal increase and 4V noise)
•
A reason to decrease the flash lamp power (<270 V, short pulse) is e.g. a signal overflow
at room temperature using a gain value of 1260 and lower -> especially at thin samples
(= low mass) or samples with a low specific heat the energy must be decreased (to
prevent the heating of sample of a few Kelvin which leads to wrong diffusivity values).
Usually, the negative influence of signal noise is smaller than the influence of too much
energy. Due to more noise a little bit higher standard deviation can be expected, but the
mean value of 5 shots is the same.