![NETZSCH LFA 447 Nanoflash Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/netzsch/lfa-447-nanoflash/lfa-447-nanoflash_operating-instructions-manual_1675675058.webp)
Operating the Instrument
LFA 447
TM
01/09 J:\LFA447\CHAPTER4.doc
7
to coat the samples:
1.
Clean all samples with a suitable solvent (including the reference sample if it has been
graphite coated previously) and lay them side by side (maintaining sample
identification).
2.
Shake the can and hold it approximately 10-12 inches from the samples.
3.
Fully depress the valve aiming to one side of the row of samples for a short time (< 1s) to
initiate the spray. Slowly sweep across the row of samples several times while
maintaining a constant distance, applying one thin coat.
4.
Wait approximately 2 minutes for the coat to be dry.
5.
Repeat steps 2-4 three more times.
6.
Turn the samples over, taking care that the coating on the sample faces will not be
damaged.
7.
Repeat steps 2-4 two or three times.
Sample Dimensions
•
Thickness
•
Diameter/Length
Sample Mass
The sample mass is used for the calculation of the density which is necessary for the
determination of the thermal conductivity.
Sample Density
The bulk density of the sample (in g/cm
3
) must be measured for the following tests:
•
specific heat
•
thermal conductivity
Содержание LFA 447 Nanoflash
Страница 2: ...01 09 J LFA447 CHAPTER1 DOC Chapter I General Information LFA 447TM...
Страница 18: ...01 09 J LFA447 CHAPTER2 DOC Chapter II Installation LFA 447TM...
Страница 25: ...01 09 J LFA447 CHAPTER3 DOC Chapter III System Components LFA 447TM...
Страница 50: ...01 09 J LFA447 CHAPTER4 doc Chapter IV Operating the Instrument LFA 447TM...
Страница 70: ...Operating the Instrument LFA 447TM 01 09 J LFA447 CHAPTER4 doc 19...
Страница 73: ...01 09 J LFA447 CHAPTER5 DOC Chapter V Appendix LFA 447TM...