background image

P a g e

 | 

17 

 

 

 

 

For uploading test results to a PC refer to “Print” section in this 
manual. 

Cardio version Test 

 

1.  From the defibrillator menu select “Card” with F1 for the 

cardio version menu as shown below. The default 
waveform is set to ECG @ 80BPM.  

2.  Set the defibrillator in synchronous mode.  

3.  For synchronizing the ECG through the ECG leads, 

connect the ECG leads of the defibrillator to the ECG 
output connectors.  The defibrillator will synchronize with 
the ECG rate and instantly discharge the energy. The 
energy and the sync time will be displayed. 

4.  To set the ECG rate and measure the sync time press 

F1. Each time F1 is pressed the rate will be increased 
by 10 BPM.  

5.  To change the amplitude of the ECG press F4 and 

select the desired amplitude. The selected amplitude will 
blink and the amplitude will update. 

6.  To save the cardio test result press save. The cardio 

version (Sync) time and the ECG rate and the amplitude 
set will be saved under the Equipment ID entered 
earlier. 

Содержание Delta 3300

Страница 1: ...Manual 660 417 Delta 3300 Quick Reference Manual Full version of the manual is included in the CD Compact Easy to Use Best Value Defibrillator and Transcutaneous Pacemaker Analyzer PN 660 417 Innovat...

Страница 2: ...3 Delta 3300 Features 3 Intended Use 4 Specifications 4 Defibrillator Analyzer 4 Transcutaneous Pacemaker 5 General Specifications 7 Part Numbers and Ordering Information 8 Preparation for Use 9 Inst...

Страница 3: ...Pulse Measurement 21 Pace Refractory Test 22 Sense Refractory Test 24 Sensitivity Test 24 Noise Immunity Test 25 ECG Simulation 26 Setup 26 Save and Print Test Results 28 Introduction 28 Saving Test R...

Страница 4: ...the Delta 3300 If any damage to the unit is observed please contact Netech Customer Service The Service Return Form may be obtained at our web site http www netechcorp us RmaRequests or from the Nete...

Страница 5: ...ocation where it might come in contact with a patient The Delta 3300 Defibrillator Analyzer is intended to be used within the published specifications Any unauthorized alteration or modifications will...

Страница 6: ...ged The DELTA 3300 samples the pulse calculates and displays the energy delivered In addition to the defibrillator function DELTA 3300 tests all types of transcutaneous pacemakers The dedicated functi...

Страница 7: ...ator Analyzer Energy Output Measurement Maximum Energy 800 J Display Resolution Based on Auto Range Auto Range 2 9 99 J 10 99 9 J 100 800 J Resolution 0 01 J 2 9 99 0 1 J 10 99 9 1 J 100 500 Accuracy...

Страница 8: ...neous Pacemaker Test Load Range 50 to 1100 ohms in step of 100 ohms Accuracy 50 500 ohm 1 500 1100 ohm 1 5 Pulse Measurements Amplitude 4 to 400 mA 50 ohm load Accuracy 1 or 1mA Max Amplitude 400 mA a...

Страница 9: ...cy 5 ms Patient Simulator Waveform Simulation ECG NSR ECG Rate 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200 220 240 260 280 300 310 BPM Accuracy 1 Amplitude 0 1 0 25 0 5 1 0 1 5 2...

Страница 10: ...litudes 1 0 mV Lead II Amplitude accuracy 2 Lead II 1 0 mV General Specifications Temperature Range Operating 15 C to 35 C Storage 0 C to 50 C Display Type LCD with LED back lighting Alphanumeric form...

Страница 11: ...er 2 660 610 6353 Standard Pacemaker Adapter cable 660 HARD CASE Hard Carrying Case 660 ProComPrint ProComPrint Software Ver1 0 Open Source Optional Accessories 661 651 HP Interface cable for HP AED M...

Страница 12: ...on Battery Charger 110 220 V input Paddle contact adapter for AEDs 2 Pieces Transcutaneous Pacer Interface cables User manual Print software CD Before getting started it is important to get familiar w...

Страница 13: ...eys are interactive associated with the respective function selections ESC Brings you back to the previous menu SAVE Saves the displayed test data in to the storage memory under the unique ID of the u...

Страница 14: ...easy identification of the leads ECG High Level Output Connectors High level ECG output connectors are marked on the side panel next to the ECG leads Power ON OFF Switch The power ON OFF switch is loc...

Страница 15: ...ed on the back of the unit It contains a 7 4 Volt Li Ion rechargeable battery specially designed for the unit with a built in protection circuit The battery is partially charged prior to shipment When...

Страница 16: ...ut a 4 digit Equipment ID of the unit under test DUT The test results that are saved will be designated under this unique ID and available for uploading to a PC and print Therefore it is very importan...

Страница 17: ...main Menu press F4 CONFG From The Configuration Menu press F1 TIME to set Hour Minute and Seconds Use F1 or F2 to set the hour Press F4 NEXT to change the minute Use F1 or F2 to set the minute Pres F...

Страница 18: ...rely place the defibrillator s APEX and STERNUM to the respective contact plates on the Delta 3300 This ensures correct signal polarity of the stored waveform through the high level output ECG leads a...

Страница 19: ...act on the defibrillator or on the defibrillator ECG leads or on a scope through the High level output jacks To view the waveform through the paddle or through the ECG leads or on a scope press F4 Wou...

Страница 20: ...ECG output connectors The defibrillator will synchronize with the ECG rate and instantly discharge the energy The energy and the sync time will be displayed 4 To set the ECG rate and measure the sync...

Страница 21: ...3 Select Start F4 4 Set the desired count down time The countdown starts from the set value and shows on the screen 5 Place the paddle firmly on the contact plates and wait for the beep 6 At the beep...

Страница 22: ...that should prompt the user to shock or no shock WARNING Before getting started follow the manufactures safety instructions and test procedures Select the function key marked AED The following screen...

Страница 23: ...ow the manufacturer s test procedures 7 The AED will analyze the waveform and audio alert for shock or no shock 8 If shock advised the AED will discharge and the discharged energy will be displayed 9...

Страница 24: ...age rate width and energy It also conducts pacemaker demand sensitivity tests measuring and displaying refractory periods and immunity tests which determine the pacemaker s susceptibility to 50 60 Hz...

Страница 25: ...sult displayed will be saved with Equipment ID set in the beginning Pace Refractory Test Refractory period of a demand pacemaker is the time interval during which a pacemaker is insensitive to ECG sig...

Страница 26: ...From the Pacer Test Main Menu select Refr refractory with F1 2 The following Refractory Menu will be displayed 3 Set the pacemaker in demand mode 4 Select Pace F1 The pace refractory and the pulse ra...

Страница 27: ...een stops the result is displayed 1 Select Sens F4 The Sense refractory and the pulse rate will be displayed after a few pacer pulse cycles and the flashing of stops and the result will be displayed 2...

Страница 28: ...to change the Waveform 4 When done select F4 to start the test 5 The rate and the sensitivity in mV will be displayed 6 If the selected width of the waveform is greater than 1 3 of the pacer width a...

Страница 29: ...nu The abbreviations used in the menu are described below NSR Normal Sinus Rhythm ARTH Arrhythmia waveforms PACR Simulated Pacer waveform PERF Performance waveforms Sine Triangle Square Setup 1 If the...

Страница 30: ...ed waveform rate and amplitude is displayed and the waveform can be observed on the monitor of the DUT 6 The selected waveform and the rates and amplitudes are saved until it is changed 7 To change th...

Страница 31: ...full it will flash an error message indicating that no more records can be saved The test results saved are ready to be uploaded to a PC using the ProComPrint software supplied with unit The ProComPri...

Страница 32: ...When the installation is complete a window appears as shown below To setup the software follow the instructions below Insert the ProCom Software CD in the PC Double click on the Setup Exe The followin...

Страница 33: ...which the results are to be saved need to be set The default folder is set to C Program Files x86 ProComPrint To change the folder click on Browse and set the folder in the desired directory Click Ne...

Страница 34: ...P a g e 31 Click Next The following screen will appear Click Install Click Finish in the next screen The following ProPrint Main screen will appear...

Страница 35: ...he RS232 or RS232 USB cable to the PC Turn the Delta 3300 Power ON A message STM32 Boot loader will appear on the screen as shown below conforming that a communication is established between the PC an...

Страница 36: ...n this will down load all the saved test results from the MultiPro memory into the top screen Select Equipment ID If multiple units are being tested the results of the individual units can be down loa...

Страница 37: ...n in the below Uploading Test Results To upload all the test results select PRN F2 from the MAIN MENU The Print MENU will be displayed The following are the Print menu options HEAD Select F1 for uploa...

Страница 38: ...P a g e 35...

Страница 39: ...ciencies 2 The Delta 3300 is returned to Netech at the buyer s expenses only after obtaining the proper RMA authorization from Netech 3 Netech will not be liable for any incidental or consequential da...

Страница 40: ...tech Products Pressure Meters Electrical Safety Analyzers AED Tester Infusion Pump Analyzer Patient Simulators EEG Simulator High Quality Best Value 2 Yr Warranty ServingBiomedicalIndustryWorldwide Si...

Отзывы: