Set desired energy on the defibrillator.
•
Discharge the defibrillator. When the unit detects the
voltage, the flashing ‘????’ on the screen stops and
the discharge energy as well as peak voltage and
current will be displayed.
•
An important feature of the Delta 3300 is the auto
ranging function, which ensures precision energy
measurements. Energy readings under 10 will be
displayed with 2 decimals, 10 joules to 100joules
one decimal, and over 100 Joules no decimal.
•
The pulse waveform is captured and saved until the
next discharge. The waveform can be viewed
through the paddle contact on the defibrillator, or on
the defibrillator ECG leads, or on a scope through
the High level output jacks.
•
To view the waveform through the paddle or through
the ECG leads or on a scope, press F4 “Wout”. The
waveform will be expanded 100 times the time base.
•
If the defibrillator is discharged again, the current
reading will be erased and the new energy reading
along with the peak voltage and current will appear
on the display. The stored waveform will be
overwritten with the new discharge pulse.
Saving Test Results
To save the test result displayed each time after discharge,
press the save key.
All saved test results will be saved in the internal flash memory,
tagged with the Equipment ID set in the earlier step. As many
as 2000 test records can be saved in the internal memory.