Front Panel
Service Manual
10030881 Rev. B 12-2008
19
SatSeconds Icon
—The SatSeconds feature provides alarm management for mild or
brief SpO
2
limit violations. When the SatSeconds feature is enabled, the SatSeconds
circle icon fills in the clockwise direction as the SatSeconds alarm management
system detects SpO
2
readings outside of the limit setting. The SatSeconds icon
empties in counterclockwise direction when SpO
2
readings are within limits. When
the SatSeconds icon reaches full, a medium priority alarm sounds. Visible in all
display views. For more information about using the feature, reference the
N-600x
Operator’s Manual
.
Fast Response Mode Icon
—Appears at bottom right of the menu bar when
enabled. The oximeter algorithm responds to changes in the SpO
2
data at differing
rates: two to four seconds in Fast Mode and five to seven seconds in Normal Mode
when calculating %SpO
2
. The response mode setting does not affect the algorithm’s
calculation of pulse rate, nor does it influence the recording of trend data, which
both occur at one-second intervals. The response mode, however, may impact the
SPD alarm behavior. Visible in all display views when enabled. For more information
about using the feature, reference the
N-600x Operator’s Manual
.
Neonate Mode Icon
—Appears at the bottom far right of the menu bar when
enabled. Visible when the alarm limits are set to neonate limit values, but not when
set to adult limit values. For more information about using the feature, reference the
N-600x Operator’s Manual
.
Monitoring Values
WARNING
Failure to cover the OxiMax pulse oximetry sensor site with opaque material in high
ambient light conditions may result in inaccurate measurements.
The OxiMax N-600x pulse oximeter continuously assesses the quality of the pulse
oximetry signal while monitoring patient SpO
2
and pulse rate. Oximeter front panel
values reflect the data derived from monitoring.
The N-600x pulse oximeter algorithm automatically extends the amount of data
required for measuring SpO
2
and pulse rate depending on the measurement
conditions.
•
Normal conditions
—During normal measurement conditions the averaging time is six
to seven seconds (approximately three seconds in Fast Mode).
•
Brief abnormal conditions
—During conditions such as those caused by low perfusion,
interference (e.g., external interference like ambient light), or a combination of these, the
oximeter automatically extends the amount of data required beyond seven seconds. If
the resulting dynamic averaging time exceeds 20 seconds, the pulse search indicator is lit
solid and %SpO
2
and Pulse Rates update every second.
•
More severe conditions
—As these conditions extend, the required amount of data
continues to increase. If the dynamic averaging time reaches 40 seconds, the pulse search
indicator flashes to denote a loss-of-pulse condition and %SpO
2
and pulse rate flash
values of zero. This activates the audible alarm state.
To Purchase, Visit
Содержание OXIMAX N-600X Series
Страница 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Страница 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Страница 24: ...General Introduction 24 10030881 Rev B 12 2008 Service Manual Figure 6 Second Event No SatSeconds Alarm...
Страница 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Страница 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Страница 160: ...Troubleshooting 160 10030881 Rev B 12 2008 Service Manual To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Страница 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Страница 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Страница 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Страница 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Страница 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Страница 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Страница 198: ...Oximeter Schematics 198 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 11 of 13...
Страница 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Страница 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Страница 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Страница 209: ......