Managing the Oximeter Backlight
Service Manual
10030881 Rev. B 12-2008
61
The movement of the blip bar, the plethysmographic waveform, or the flashing heart
icon indicates real-time data display. The pulse beep tone is an audible indicator of
the real-time patient data.
Caution
If the pulse beep tone does not sound with each pulse, the pulse beep volume is set to zero,
the speaker is malfunctioning, or the signal is corrupt.
When first applying an OxiMax pulse oximetry sensor to a patient, the oximeter may
lose a pulse signal. If a pulse signal is lost, an alarm sounds and a poor signal
condition message displays on the oximeter screen. At this point, the oximeter
displays [--- / ---](three dashes over three dashes) and remains in the Pulse Search
mode for five seconds before displaying the poor signal condition screen. The poor
signal condition screen is part of the Sensor Messages feature.
Figure 21.
Poor Signal Condition Message
Sensor Detection Failure
Upon successful completion of the POST process, the oximeter sounds a one-second
tone indicating that the oximeter has passed POST.
Should the oximeter fail to detect an OxiMax pulse oximetry sensor, the oximeter
displays dashes [ - - - ] and the Pulse Search indicator does not light.
Managing the Oximeter Backlight
Adjust the brightness, contrast, and backlight of the oximeter screen to suit each
individual situation.
To turn off the oximeter backlight
1. Press the LIGHT softkey.
2. Then press OFF.
Note:
Any of the following conditions turn on the backlight:
•
Pressing any of the softkeys
•
Pressing and holding the HELP/CONTRAST key
•
Pressing the ALARM SILENCE key
•
Any alarm
To adjust the backlight brightness
1. With the oximeter in the normal monitoring mode, press the LIGHT softkey.
To Purchase, Visit
Содержание OXIMAX N-600X Series
Страница 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Страница 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Страница 24: ...General Introduction 24 10030881 Rev B 12 2008 Service Manual Figure 6 Second Event No SatSeconds Alarm...
Страница 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Страница 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Страница 160: ...Troubleshooting 160 10030881 Rev B 12 2008 Service Manual To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Страница 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Страница 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Страница 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Страница 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Страница 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Страница 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Страница 198: ...Oximeter Schematics 198 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 11 of 13...
Страница 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Страница 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Страница 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Страница 209: ......