Operating the Oximeter
64
10030881 Rev. B 12-2008
Service Manual
1. With the oximeter in the normal monitoring mode, press the ALARM SILENCE key
until the alarm silence duration setting displays.
2. While pressing the ALARM SILENCE key, press and hold the ADJUST UP key until
OFF displays. Release both keys.
Note:
Once audible alarms are disabled, the orange LED above the ALARM SILENCE key
lights to indicate the disabled alarm state.
To select the standby mode
Normally, use the standby mode setting for the patient who must temporarily leave
the oximeter.
The standby mode enables the oximeter to retain alarm limit settings in effect when
a caregiver temporarily removes a patient’s sensor.
1. Verify the oximeter alarm limits are configured to the monitored patient.
2. Disconnect the sensor from the oximeter.
3. Press the ALARM SILENCE key to silence the audible alarms.
4. Press the ALARM SILENCE key to disable the alarm messages.
Figure 24.
Standby Mode
The oximeter is now in the standby mode. Reconnect the sensor to the oximeter and
the patient to return to normal monitoring.
Using Oximeter Softkey Menus
Navigating Menu Options
Hierarchy
The oximeter’s softkey menu hierarchy allows the user or technician to configure and
operate the oximeter, select options, and view trend or event data.
Note:
The oximeter compiles trend data on each entry or re-entry to the Trend menu. To refresh
active trend data, re-enter the Trend menu from the Main menu.
Sensor sub-menu choices differ, depending on what type of patient alarm event data
are stored in the sensor chip.
To Purchase, Visit
Содержание OXIMAX N-600X Series
Страница 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Страница 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Страница 24: ...General Introduction 24 10030881 Rev B 12 2008 Service Manual Figure 6 Second Event No SatSeconds Alarm...
Страница 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Страница 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Страница 160: ...Troubleshooting 160 10030881 Rev B 12 2008 Service Manual To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Страница 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Страница 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Страница 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Страница 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Страница 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Страница 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Страница 198: ...Oximeter Schematics 198 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 11 of 13...
Страница 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Страница 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Страница 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Страница 209: ......