Troubleshooting
146
10030881 Rev. B 12-2008
Service Manual
Error Codes for Non-correctable Failures
Contact Nellcor’s Technical Services Department, or your local Nellcor representative
should the oximeter detect a non-correctable failure. When one of the listed errors
occurs, several events also occur.
•
The oximeter sounds a low priority alarm that cannot be silenced except by power
cycling the oximeter.
•
Measurements stop.
•
All information on the screen vanishes, displaying an error code message. Error code
messages begin with EEE and are followed by an error code.
Note:
Cycling the power may clear the displayed error code.
Table 26.
Error Code Listing for Non-correctable Failures
Error Code
Category
Description
1
%SpO
2
front end
%SpO
2
front end RAM error.
2
%SpO
2
front end ROM/code integrity error.
3
%SpO
2
front end reported a bad Cyclic Redundancy Check (CRC).
4
%SpO
2
front end reported FSP message not allowed.
5
%SpO
2
front end reported illegal value sent in FSP message %SpO
2
front
end.
6
%SpO
2
front end reports calibration (offset) failure.
9
%SpO
2
front end reported syntax error in FSP message.
10
Over-current limit in %SpO
2
front end has tripped.
11
%SpO
2
front end reports incorrect system voltage.
12
%SpO
2
front end reports other hardware problem.
14
%SpO
2
front end reports communication channel overflow.
16
%SpO
2
front end reports watch dog time out.
17
%SpO
2
front end
%SpO
2
front end reports that sensor appears defective.
18
%SpO
2
front end reports internal register appears modified from
expected value.
19
%SpO
2
front end reports signal out-of-range.
48
%SpO
2
front end reports spurious interrupt.
49
%SpO
2
front end reports internal buffer overflow.
50
%SpO
2
front end reports intermittent error.
51
%SpO
2
front end reports digital communications error.
52
%SpO
2
front end reports warmer error.
53
Front end data not received.
Содержание OXIMAX N-600X Series
Страница 8: ...viii 10030881 Rev B 12 2008 Service Manual...
Страница 12: ...Safety Information 12 10030881 Rev B 12 2008 Service Manual...
Страница 24: ...General Introduction 24 10030881 Rev B 12 2008 Service Manual Figure 6 Second Event No SatSeconds Alarm...
Страница 44: ...Theory of Operations 44 10030881 Rev B 12 2008 Service Manual...
Страница 96: ...Managing the Data Port 96 10030881 Rev B 12 2008 Service Manual...
Страница 160: ...Troubleshooting 160 10030881 Rev B 12 2008 Service Manual To Purchase Visit Avobus com or call 1 800 674 3655...
Страница 188: ...Oximeter Schematics 188 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 1 of 13...
Страница 189: ...Service Manual 10030881 Rev B 12 2008 189 Main PCB Schematic Sheet 2 of 13...
Страница 191: ...Service Manual 10030881 Rev B 12 2008 191 Main PCB Schematic Sheet 4 of 13...
Страница 192: ...Oximeter Schematics 192 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 5 of 13...
Страница 194: ...Oximeter Schematics 194 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 7 of 13...
Страница 195: ...Service Manual 10030881 Rev B 12 2008 195 Main PCB Schematic Sheet 8 of 13...
Страница 197: ...Service Manual 10030881 Rev B 12 2008 197 Main PCB Schematic Sheet 10 of 13...
Страница 198: ...Oximeter Schematics 198 10030881 Rev B 12 2008 Service Manual Main PCB Schematic Sheet 11 of 13...
Страница 200: ...Service Manual 10030881 Rev B 12 2008 200 Main PCB Schematic Sheet 13 of 13...
Страница 201: ...Service Manual 10030881 Rev B 12 2008 201 Main PCB Assembly Drawing Front View...
Страница 204: ...Oximeter Schematics 204 10030881 Rev B 12 2008 Service Manual...
Страница 209: ......