Technical Information
N-560 Operator’s Manual
89
Figure 6: Alarm Response with
SatSeconds
Saturation levels may fluctuate rather than remaining steady for a period
of several seconds. Often, %SpO
2
levels may fluctuate above and below
the alarm limit, reentering the non
-
alarm range several times.
During such fluctuation, the N
-
560 integrates the number of %SpO
2
points, both positive and negative, until either the
SatSeconds
limit
(
SatSeconds
setting) is reached, or the %SpO
2
level returns within a
normal range and remains there.
SatSeconds
“Safety Net”
The
SatSeconds
“Safety Net” is for patients whose saturation levels
frequently go below the limit but do not stay below the limit long enough
for the
SatSeconds
setting to be reached. When 3 or more limit violations
occur in 60 seconds, an alarm will sound even if the
SatSeconds
setting
has not been reached.
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Страница 12: ...Introduction 4...
Страница 22: ...Description of Controls Indicators and Symbols 14...
Страница 50: ...Using the N 560 42...
Страница 54: ...N 560 Trend 46...
Страница 70: ...Printing 62...
Страница 76: ...Sensors and Accessories 68...
Страница 82: ...Operator s Menu 74...
Страница 90: ...Troubleshooting 82...
Страница 98: ...Technical Information 90...
Страница 118: ...Specifications 110...
Страница 122: ...Index 114...
Страница 123: ......